June 2022 – TD 320 OPERATING MANUAL – TURBIDITY SENSOR 4296
Page 65
4.8.1 Mandatory
In the
Mandatory
section you will find only one property,
Node Description
. All sensors are given
a
Node Description
text like Turbidity Sensor #xxx (xxx is the serial number of the sensor) by
default. The user can modify this
Node Description
text if required. Be aware that the
Node
Description
changes to
*Corrupt Configuration
if it has lost the configuration in flash. Contact
the factory if this happens. The configuration is saved in two sectors in flash memory. A flash
sector can be corrupted if the power is lost during the saving of new configuration. The double
flash sector saving ensures that it does not lose the configuration. If one of the sectors is
corrupted, the other sector is used and saved to the corrupt sector.
4.8.2 Site Info
In the
Site Info
section, you will find only one property.
Owner
is optional information to be
entered to store information about the owner. This setting is not used in calculation. By default,
this setting is empty.
4.8.3 Calibration
Figure 4-22: Mandatory in User Maintenance
Figure 4-23; Site Info in User Maintenance