Chapter 3 Controller Descriptions and Operation
3
04 00210 0 28.00 85.0 000
Keep the temperature at 28.00 °C and humidity at
85.0 %RH for 210 min.
05 00060 0 10.00 0.0 000
Change the setting in 60 min for temperature from
28.00 °C to 10.00 °C, with linear rate 0.3 °C/min. No
humidity required
06 00420 0 10.00 0.0 000
Keep the temperature at 10.00 °C for 420 min, no
humidity required.
07 00060 0 25.00 75.0 000
Change the setting in 60 min for temperature from
10.00 °C to 25.00 °C, with linear rate 0.3 °C/min, and
humidity from a random value to 75.0 %RH in a linear rate
according to the humidity difference and in 60 min.
08 00420 0 25.00 75.0 000
Keep the temperature at 25.00 °C and humidity at
75.0 %RH for 420 min.
Compare example 1 and example 2 for ease of understanding.
Fig 24
At screen of Program editing, other functional keys as follows:
[INSERT]: Allow an inclusion of a program segment between two existing program segments for
editing. Select a segment number, and the new segment will be inserted before the selection
with all data of 0 for editing. The selected segment and all thereafter increase a segment
number automatically.
[DELETE]: Delete a selected program segment for removal. The selected segment and all
thereafter decrease a segment number automatically.
[COPY]: Copy a selected source program segment to target program segment.
[<], [>]: Move up and down between Segments, or move among TIME, WAIT, TEMP, HUMD and
LIGHT.