XTEMP ENVIRONMENTS (SHANGHAI) CO., LTD.
http://www.xtemp-usa.com/
15
/
55
DSEO0711-2-V211119/UXP8401-PG.CC.GE.AR-CD7003/5~+45
℃
Chapter 2 Installing and Operating
2
2.2.2 Back view
Fig 2 Back view
04 Exhausting-air outlets
06 Hatch/Access port
22 Power cable
25 Fixed casters
30 Connection for exhaust opening
31 Condensate outlet
57 Water inlet for humidifier
70 Humidifier control module
213 Ethernet interface (optional)
Perform regular check or maintenance for the following at designated internals:
Clean the cabinet interior weekly or where necessary!
Decontaminate cabinet interior and wire shovels on case-by-case bases if necessary!
Check the air tightness of the door gasket at three months internals or the first time application
after storage of service!
Check and clean the condenser at six months internals or where necessary!
Caution: Hazard of strong optical radiation!
Lamp canopies incorporated either at both sides, or within the cabinets generate strong visual
radiation which could cause a hazard. Use special care when trying to get access to the cabinet
interior:
a)
Use glass ware or equivalent PPE as appropriate;
b)
Check functionality of dimming capability of the lamp canopies once door opened, and
confirm that the light intensity within safe limit;
c)
Access to the side lamp canopies only the well-trained personnel and with appropriate
PPE against optical radiation, electric shock and burn hazard
;
Never disable the dimming sensor for safe light intensity.