DSEO0711-2-V211119/UXP8401-PG.CC.GE.AR-CD7003/5~+45
℃
Chapter 2 Installing and Operating
2
2.2 Descriptions
2.2.1 Front view
Fig 1 Front view
01 Power SW with lock
02 HMI touch screen
13 Lamp canopy
(
CC/GE
)
14 Perforated shelves
17 Interior glass
18 Front door
19 The levelers
34 Ventilator
35 Glass door alarm switch
42 Temperature and humidity aspirator
58 Sockets for PAR and Lux transmitter
71 Ultrasonic humidifier
72 USB socket (Optional)
123 Drip pan
145 Cable socket for portable T/H aspirator
150 Connection for air sampling tube
226 Proximity switch for dimming of lamp canopy
The first time powering up after repairing or standstill for long period, or at designated intervals
under normal use, check the functionality of the following parts or components:
Check the air tightness of the door seal at six months internals!
Check and clean the condenser at six months internals or where necessary!
Clean the cabinet interior weekly or where necessary!
Decontaminate cabinet interior and wire shovels on case-by-case bases if necessary!
Keep clear of inlets and outlets of back plenum for the circulating air. Excessive hoarding of
specimen or blockage of the air circulating may cause frost accumulation on surface of the
evaporator, localized temperature distribution and degraded performance!