90
2000I & 2000M User Guide, Rev B
Table 9. Thermocouple Simulation & Measurement Specifications
Type
Simulation
Resolution
Measurement
Resolution
Measurement Range
B
0.01ºC
0.01ºC
0.15ºC
0.08ºC
+200 to +300ºC
+300 to +1820ºC
E
0.01ºC
0.01ºC
0.01ºC
0.15ºC
0.05ºC
0.01ºC
-270 to -260ºC
-260 to -70ºC
-70 to +1000ºC
J
0.01ºC
0.01ºC
0.02ºC
0.01ºC
-210 to -140ºC
-140 to +760ºC
K
0.01ºC
0.01ºC
-140 to +1372ºC
N
0.01ºC
0.03ºC
-200 to +1300ºC
R, S
0.01ºC
0.01ºC
0.06ºC
0.03ºC
0 to +200ºC
+200 to +1768ºC
T
0.01ºC
0.01ºC
0.02ºC
0.01ºC
-200 to +100ºC
+100 to +400ºC
Note:
For tables 10a and 10b, the +3ppm + 1
µ
V absolute accuracy adder for the
Xitron Technologies production calibration may be substituted by the
accuracy figures of your calibration equipment, changing the absolute
accuracies provided.