88
2000I & 2000M User Guide, Rev B
Note:
To maintain accuracy whenever outside this temperature range, apply
the relevant temperature coefficient adder.
Settling Time: -------Less than 2 seconds to within 10ppm of 0.25µV
Drive Capability: ---15mA max. (50
Ω
output impedance below 220mV, 0.1
Ω
above 220mV)
Overload:-------------Output fully protected against short-circuits
Long-Term Effects: Add 25ppm of output to 1 year specification for 3 year
accuracy
DC Current Output Specifications
The accuracy figures given in Table 8 are valid throughout the temperature
range of 18º to 28ºC. The 24 hour stability figures are for constant load and
temperature (within 1ºC) and are measured using a DC to 2Hz bandwidth.
The accuracy figures are relative to calibration standards. Add 15ppm + 20pA
to the current figures for absolute accuracy.
Table 8: DC Current Output
Levels
< 22µA
22 – 220µA 0.22 – 2.2mA
2.2 - 22mA
Resolution
10pA
100pA
1nA
10nA
24 Hr. Stability (ppm + nA)
7.0 + 0.03
5.0 + 0.15
5.0 + 1.0
5.0 + 15
90 Day Accuracy (ppm + nA)
50 + 0.05
40 + 0.15
30 + 1.0
40 + 20
1 Year Accuracy (ppm + nA)
70 + 0.1
60 + 0.25
50 + 2.0
60 + 30
Temperature Coefficient P
(ppm + nA) /
°
C
5.0 + 0.015
5.0 + 0.1
5.0 + 1.0
5.0+1.0
Pk-PK Noise 0.01 – 10Hz
(ppm+ nA)
1.5 + 0.02
1.2 + 0.1
0.8 + 0.5
0.8 + 5.0
RMS Noise 10Hz – 10kHz
200pA
750pA
5nA
50nA
DC current output is available in both the 2000I and the 2000M instruments.
Summary of Contents for 2000I
Page 1: ...USER GUIDE 2000I 2000M DC Power Source Instruments...
Page 2: ......
Page 10: ...10 2000I 2000M User Guide Rev B...
Page 20: ...20 2000I 2000M User Guide Rev B...
Page 44: ...44 2000I 2000M User Guide Rev B...
Page 76: ...76 2000I 2000M User Guide Rev B...
Page 96: ...2000I 2000M Instrument s User Guide Revision B 96...