![XITRON 2000I User Manual Download Page 82](http://html2.mh-extra.com/html/xitron/2000i/2000i_user-manual_5005162082.webp)
82
2000I & 2000M User Guide, Rev B
Note:
Command types other than those listed in Tables 4, 5 or 6 are always
discarded and ignored.
Table 4: Command Characters
Front
Panel Key
Characters
Received
Front
Panel Key
Command
Character
Received
0
0
. (period)
. (Dec. Pt.)
1
1
+/–
–
2
2
V, Volts, or ENTER
V
3
3
A, Amps, or ENTER
A
4
4
m (milli-)
m
5
5
µ (micro-)
u
6
6
ºC, or ENTER
C
7
7
ºF, or ENTER
F
8
8
CLR
E
9
9
MNU
M
When using the IE-BAT interface, each character may be sent with the
EOI
line asserted, in which case the display will show the progress of the
command that is being built in this manner.
Table 5: Shortcuts to Command Character Sequences
Char/Seq
Description of Action
@
Initiates the output of the display data (RS-BAT only)
D
Selects the “Adjust” mode.
I
Commands an internal calibration sequence.
J0
Selects “Passive CJ” (i.e., turns off internal Active CJC)
(2000M)
J1
Selects “Active CJC” activity
(i.e., turns on internal Active CJC). (2000M)
K0
Deselects the “Calculated Output” mode of operation.
K1
Selects “Calculated Output” mode of operation, using the
presently set pair of user defined points and units.
N
Deselects the “Adjust” mode.
Summary of Contents for 2000I
Page 1: ...USER GUIDE 2000I 2000M DC Power Source Instruments...
Page 2: ......
Page 10: ...10 2000I 2000M User Guide Rev B...
Page 20: ...20 2000I 2000M User Guide Rev B...
Page 44: ...44 2000I 2000M User Guide Rev B...
Page 76: ...76 2000I 2000M User Guide Rev B...
Page 96: ...2000I 2000M Instrument s User Guide Revision B 96...