User Manual
Wildcat
Company confidential.
This document is the property of Xenics. It may not be reproduced
–
completely or partially
–
or passed to a third
party without written permission from Xenics.
Xenics nv
T +32 16 38 99 00
Doc Ref: ENG-2020-UMN001
Ambachtenlaan 44
F +32 16 38 99 01
Issue: R001
BE-3001 Leuven
Belgium
www.xenics.com
Date:
XF-104_03/20-12-2012
Page 38 of 58
Digital output
Both digital outputs can be operated in parallel and offer the same functionality.
Digital output control
– Source
defines the signal type placed on the digital output.
•
Disable
disables the output channel.
•
High
drives the output to a high
(≥ 2.9 V) state
•
Low
drives the output to a
low (≤ 0.4 V) state.
•
Start of frame
makes this a trigger output and generates a pulse on the output at
the start of integration.
The digital outputs can be inverted with the
Invert
setting.
The digital outputs can be delayed by writing a time in µs higher than 0 into
Delay
.
When
Start of frame
is selected as signal source, the pulse duration in µs can be defined
by entering a value in
Pulse width.
The actual current value on the digital output can be read from
Status
, with a value of
True
if the output is
≥ 2.9 V and a value of
False
if the output is
≤ 0.4 V.