e X a c t A u t o - S c a n S y s t e m
21
eXact Auto-Scan Vacuum/Foot Switch Interface Wiring
Pin
Name
Description
1
Output
External Control Line: provides up to 120mA current sink
during measurement sequence.
2
Input
External Trigger Line: turns on laser when pulled to ground, if
laser is already on a measurement will begin.
3
+12v
Pr12 VDC, up to 500mA (not fused).
4
GND
DC Ground
INPUT
OUTPUT
The CPC1030 has a maximum
current of 120mA.