e X a c t A u t o - S c a n S y s t e m
15
7.
When finished, flip the target window back to the scan position and reattach the instrument to the
scanning trolley.
Calibration
IMPORTANT:
The calibration plaque is dramatically affected by smudge marks and dust, and must be
kept clean. It is recommended that the instrument is cleaned regularly. Refer to the cleaning procedure
in the Appendix.
1.
When prompted for a calibration, remove the eXact instrument from the scanning trolley if
required.
2.
Open the instrument (not locked) and tap the START button on the eXact screen. The instrument
will display a countdown screen and then perform the calibration.