Oxi 3310 IDS (Oxi 3315)
Dissolved oxygen
ba75944e02
10/2014
25
the salinity of the measured medium first.
When the salinity correction is switched on, the [Sal] indicator is displayed in
the measuring screen.
Stability control
(AutoRead)
& HOLD function
The stability control function (
AutoRead
) continually checks the stability of the
measurement signal. The stability has a considerable impact on the reproduc-
ibility of measured values.
The measured parameter flashes on the display
as soon as the measured value is outside the stability range
when the automatic
Stability control
is switched off.
Irrespective of the setting for automatic
Stability control
(see section 6.2.3
, page 35) in the
System
menu, you can start a
measurement with
Stability control
manually at any time.
You can switch the salinity correction on or off and enter the salinity
in the menu for calibration and measurement settings (see section
6.1.1 S
1.
Freeze the measured value with
<AR>
.
The [HOLD] status indicator is displayed.
The HOLD function is active.
You can terminate the
Stability control
function and the HOLD func-
tion with
<AR>
or
<M>
at any time.
2.
Using
<ENTER>
, activate the
Stability control
function manually.
The [AR] status indicator appears while the measured value is assessed
as not stable. A progress bar is displayed and the display of the mea-
sured parameter flashes.
The [HOLD][AR] status indicator appears as soon as a stable measured
value is recognized. The progress bar disappears and the display of the
measured parameter stops flashing.
The current measurement data are output to the interface. Measurement
data meeting the stability control criterion are marked by AR.
You can prematurely terminate the
Stability control
function manu-
ally with
<ENTER>
at any time. If the
Stability control
function is
prematurely terminated, the current measurement data are output
to the interface without the AutoRead info.