Dissolved oxygen
Oxi 3310 IDS (Oxi 3315)
32
ba75944e02
10/2014
ibration. The evaluation appears on the display and in the calibration record.
FDO
®
925 calibra-
tion evaluation
Calibration record
(
USB output
)
Display
Calibration record
Relative slope
+++
S = 0.94 ... 1.06
++
S = 0.92 ... 0.94
or
S = 1.06 ... 1.08
+
S = 0.90 ... 0.92
or
S = 1.08 ... 1.10
Error
Error
S < 0.90
or
S > 1.10
Error elimination (see section 10 W
CALIBRATION Ox
Calibration date
01.02.2014 07:43:33
FDO 925
Ser. no. 10146858
SC-FDO 925
10158765
Relative slope
0.98
Sensor
+++