Oxi 3310 IDS (Oxi 3315)
Settings
ba75944e02
10/2014
33
6
Settings
6.1
D.O. measurement settings
6.1.1 Settings for D.O. sensors
(menu for measurement and calibration settings)
Settings
The settings are available in the menu for measurement and calibration set-
tings. To open the settings, display the required measured parameter in the
measured value display and press the
<ENTER>
key. After completing the set-
tings, switch to the measured value display with
<M>
.
Default settings are printed in
bold
.
Menu item
Possible
setting
Explanation
Calibration
/
Calibration record
-
Displays the calibration record of the
last calibration.
Calibration
/
Calibration data
storage
/
Display
-
Displays the last calibration records
(max. 10)
Calibration
/
Calibration data
storage
/
Output to
RS232/USB
-
Outputs the calibration data memory
to the interface
Calibration
/
Calibration interval
1 ...
180
...
999 d
Calibration interval
for the D.O. sen-
sor (in days).
The meter reminds you to calibrate
regularly by the flashing sensor sym-
bol in the measuring screen.
Calibration
/
Comparison meas.
on
off
Enables to adjust the measured
value with the aid of a comparison
measurement, e.g. Winkler titration.
For details, see section 5.3 C
FDO Check
/
Start FDO Check
-
Starts the FDO
®
check procedure
FDO Check
/
Check
interval
1 ...
60
...
999 d
Interval for the
FDO Check
(in days).
The meter reminds you to check the
sensor regularly by
FDO Check
sta-
tus indicator in the measuring
screen.
Sal correction
on
off
Manual salt content correction for
concentration measurements.