Operation
System 182
4 - 10
ba75504e07
04/2012
Recommended
proceeding for cleaning,
maintenance and repair
Automatic switch-on of
the maintenance
condition
The maintenance condition for IQ sensors is automatically activated
during calibration. After calibration the IQ sensor remains in the
maintenance condition until the maintenance condition is switched
off manually (see section 4.5)
during a compressed-air cleaning cycle. After the cleaning cycle is
finished the maintenance condition is automatically switched off
again.
1
Switch on the maintenance condition for the IQ sensor. The
measured value display flashes.
2
Pull the sensor out of the sample.
3
Carry out the cleaning, maintenance or repair (removal and
replacement) of the sensor.
4
Submerse the sensor in the sample again.
5
Wait until the measured value does no longer change.
6
Switch off the maintenance condition for the sensor. The
display of the sensor in the measured value display does no
longer flash.
Summary of Contents for DIQ/S 182-MOD
Page 6: ...Contents System 182 0 4 ba75504e07 04 2012...
Page 10: ...Overview System 182 1 4 ba75504e07 04 2012...
Page 14: ...Safety instructions System 182 2 4 ba75504e07 04 2012...
Page 78: ...Operation System 182 4 18 ba75504e07 04 2012...
Page 84: ...The SETTINGS menu System 182 5 6 ba75504e07 04 2012...
Page 110: ...Relay outputs System 182 6 26 ba75504e07 04 2012...
Page 126: ...Maintenance and cleaning System 182 8 2 ba75504e07 04 2012...
Page 132: ...What to do if System 182 9 6 ba75504e07 04 2012...
Page 144: ...Accessories and options System 182 11 2 ba75504e07 04 2012...
Page 148: ...Appendix store separately if required System 182 13 2 ba75504e07 04 2012...
Page 149: ......