System 182
Modbus connection
7 - 13
ba75504e07
04/2012
12
42h
Contents of register 13
(HI)
= main measured
value
429E46C2h (Float 32)
-->
79,1382
Measured parameter and
unit,
see byte 10 (measuring
mode)
13
9Eh
Contents of register 13
(LO)
= main measured
value
14
46h
Contents of register 14
(HI)
= main measured
value
15
C2h
Contents of register 14
(LO)
= main measured
value
16
00h
Contents of register 15
(HI)
= secondary measured
value
00000000h (Float 32)
-->
0
but measured value
invalid (MISSING),
see Byte 11 (measured
value status)
17
00h
Contents of register 15
(LO)
= secondary measured
value
18
00h
Contents of register 16
(HI)
= secondary measured
value
19
00h
Contents of register 16
(LO)
= secondary measured
value
20
23h
CRC (HI)
Checksum (CRC)
21
5Eh
CRC (LO)
Modbus response
Byte
Value
Information
Meaning of the value
Summary of Contents for DIQ/S 182-MOD
Page 6: ...Contents System 182 0 4 ba75504e07 04 2012...
Page 10: ...Overview System 182 1 4 ba75504e07 04 2012...
Page 14: ...Safety instructions System 182 2 4 ba75504e07 04 2012...
Page 78: ...Operation System 182 4 18 ba75504e07 04 2012...
Page 84: ...The SETTINGS menu System 182 5 6 ba75504e07 04 2012...
Page 110: ...Relay outputs System 182 6 26 ba75504e07 04 2012...
Page 126: ...Maintenance and cleaning System 182 8 2 ba75504e07 04 2012...
Page 132: ...What to do if System 182 9 6 ba75504e07 04 2012...
Page 144: ...Accessories and options System 182 11 2 ba75504e07 04 2012...
Page 148: ...Appendix store separately if required System 182 13 2 ba75504e07 04 2012...
Page 149: ......