APOLLO 4000
HYDROGEN PEROXIDE DETECTION
World Precision instruments
e-3
note: the calibration should be carried out at the temperature at which the
samples of H
o
are to be measured. this can be accomplished by placing the
vial and stand in a water bath at the appropriate temperature, and allowing the
temperature of the solution in the bottle to equilibrate with the water bath.
5.
once the current has achieved a stable value. it is ready for calibration.
Creating a calibration curve
to create a calibration curve, the user measures
the current (pA) generated by the addition of
different amounts of H
o
to the calibration
solution.
1.
After the current is stable, zero the baseline.
2.
Add a known volume of the H
o
standard
solution to the PBs buffer solution.
For example, add 0, 5, 0, 0, 40, 80ml H
o
standard solution (.0mm) into 0ml PBs buffer solution.
the output from the H
o
will like similar to the example shown in
Fig. E1
.
From this output, a calibration curve (
Fig. E2
) can
be created by plotting the changes in current (pA)
against the changes in concentration (mm). the
slope of this curve indicates sensitivity of the probe.
-100
100
300
500
700
900
1100
1300
-30
20
70
120
170
time (sec)
current (pA)
0.5
µ
M
1
µ
M
2
µ
M
4
µ
M
8
µ
M
Figure E1
— H
2
O
2
calibration
output.
y = 73.296x + 6.4857
R
2
= 0.9988
0
100
200
300
400
500
600
700
0
2
4
6
8
10
Concentration of H
2
O
2
(
µ
M)
Current (pA)
Figure E2
—
ISO-HPO-2
calibration curve
based on Figure 1
Summary of Contents for APOLLO 4000
Page 2: ......
Page 4: ......
Page 10: ...APOLLO 4000 INTRODUCTION WORLD PRECISION INSTRUMENTS A 6...
Page 48: ...APOLLO 4000 NITRIC OXIDE DETECTION WORLD PRECISION INSTRUMENTS C 24...
Page 70: ...APOLLO 4000 TROUBLESHOOTING WORLD PRECISION INSTRUMENTS F 2...
Page 72: ......
Page 74: ......