APOLLO 4000
OPERATING INSTRUCTIONS
WORLD PRECISION INSTRUMENTS
B-8
Fig. B11
• Poise Voltage
. This submenu looks (
Fig. B10
) and works similarly to the
Range control. Threre are four choices:
Nitric Oxide
—
Automatically configures
the poise voltage (
i.e.,
865 mV) on the selected
channel to measure nitric
oxide.
Oxygen
— Automatically
configures the poise
voltage (
i.e., 700 mV) on
the selected channel to
measure oxygen.
Hydrogen Peroxide
—
Automatically configures
the poise voltage (
i.e.,
400 mV) on the selected channel to measure
hydrogen peroxide.
Custom
— Allows the user to manually set
the poise voltage (
i.e., from 0-2000 mV)
(
Fig. B11
).
Selecting the wrong poise
voltage will drastically change the results
of an experiment and may render any data
invalid.
Hence the program asks for the
confirmation (
Fig. B12
) prior to changing any
poise voltage.
Fig. B10
Fig. B12