APOLLO 4000
HYDROGEN PEROXIDE DETECTION
World Precision instruments
e-
10.
suspend the tip of the new assembled probe in PBs buffer solution.
11.
After 0-5 minutes the current should no longer be offscale and will
gradually decrease with time. it may take several hours for the sensor
current to reach a low stable value, at which time it will be ready for use.
12.
if the current observed, after a few minutes in the saline solution,
increases dramatically or is offscale, then the membrane integrity is not
good and a new membrane will have to be fitted.
Storage
store the electrode with its tip immersed in electrolyte solution in the sealed vial
provided with the electrode.
.