APOLLO 4000
OXYGEN DETECTION
World Precision instruments
d-4
the physical interpretation of the
% oxygen is the percentage of
atmospheric pressure that the oxygen
present exerts. For example, in a
00% oxygen environment a reading
of 00 means that the partial pressure
of oxygen is atm (760 mm Hg). A
reading of 2 means that the partial
pressure of oxygen is 0.2 atm (60
mm Hg).
Fig. D2
shows a record of the
calibration procedure described
above. Before point the record
displays the background current
of the oxygen sensor in air. this
current value may vary from sensor
to sensor. At point the sensor is
exposed to 0 % oxygen. At point 2
the background is zeroed using the software. At point 3 and point 4 the sensor is
exposed to air (2% oxygen) and 00% oxygen, correspondingly. By using the
current at 0% (0 nm) and 00% (236 nA), a two-point calibration curve is built and
the slope is determined at 2.36 nA/%. Alternatively, a three-point calibration can be
implemented by adding the information for the current of the sensor in air and by
using linear regression software.
to present the data in % oxygen you may use the
Setup/Unit conversion
option
from the menu. A new unit for the oxygen-measuring channel must be created by
setting up 2360 pA to correspond to % oxygen (
Fig. D3
). switch to % unit as
shown in
Fig. D4
and all recorded data as well as incoming data will be displayed
in % oxygen unit.
Note:
After the calibration the trace should not be re-zeroed as this will invalidate
the calibration.
Fig. D3
—
Creating a new unit
(% oxygen).
Summary of Contents for APOLLO 4000
Page 2: ......
Page 4: ......
Page 10: ...APOLLO 4000 INTRODUCTION WORLD PRECISION INSTRUMENTS A 6...
Page 48: ...APOLLO 4000 NITRIC OXIDE DETECTION WORLD PRECISION INSTRUMENTS C 24...
Page 70: ...APOLLO 4000 TROUBLESHOOTING WORLD PRECISION INSTRUMENTS F 2...
Page 72: ......
Page 74: ......