Table 9.2: Lograte and max possible measuring time ant 4680 measurements
Log rate
Max possible measuring time
30 s
39 h
1 min
3.25 days
3 min
9.75 days
10 min.
32.50 days
30 min
13 weeks
1 h
27 weeks
3 h
19 month
4 h
approx. 2 years (with good dry batteries )
When the C I/O key is pressed in logger mode the DC 100
PRO
solely outputs
the text “Log” followed by the current measured values and then the remai-
ning memory. The cursor is locked for other commands. Logger mode is en-
ded when the device is switched OFF at the C I/O button (pressed for three
seconds). Data are transferred to a PC as explained under Section 9.1.
9.3 PrintLog/IR -> Printing out measurement records:
The records of a logger measurement can be printed out. If it is a „4-Pa-
measurement“, this will be indicated on the hardcopy.
9.4 Log/IR -> Printing out measurement reports:
Log/IR —> PRINT –The records of the last volume measurements can be
printed out directly on the TD 600 quick printer (order no. 4130), also at a la-
ter time. They are retained even after the device has been switched OFF but
are overwritten by a new pressure loss measurement or logger recordings.
The text Print appears on the display during the printing process.
9. Logger/data transfer