OWNERS OPERATING & MAINTENANCE MANUAL
PHLEBOTOMY COLLECTION
7.
Remove/pull out Arm Tube (
FIG 8-J
) and re-install in opposite end following picture on label (
FIG 8-K
)
lining up A-A or B-B based on Patient LH or RH
8.
Reinstall Leveling Screw (
FIG 8-H
), Safety Cover with 4 Screws (
FIG 9-M & P
, Flat Safety Cover
(
FIG 9-N
), & O-Ring (
FIG 9-Q
) previously removed (
Steps 2-6
)
9.
L-Arm can now be installed into the Short Arm
on opposite side (
see page 37
)
CHANGING THE POSITION OF L-ARM - CONTINUED
FAILURE TO RE-INSTALL THE
SAFETY COVER USING THE
FOUR (4) SCREWS & THE
FLAT SAFETY COVER WILL
ALLOW A PINCH POINT ACCESSIBLE BY THE
PATIENT, THE INDIVIDUAL ACCOMPANYING
THE PATIENT, AND / OR THE CAREGIVER.
Fig 8
Fig 9
2580