The new GPIB address will appear on the
display.
Step 3
Press
Line Term
to select the correct
GPIB data delimiter.
The CW generator is now configured for
360B tracking receiver mode operation.
Initiating
360B Track-
ing Mode
Turn on the 360B VNA and configure it for tracking
receiver mode operation. (Refer to the 360B VNA op-
eration manual.) Once configured, the 360B should
take control of the CW generator.
When the 360B takes control, the display of all pa-
rameters on the 680XXB is disabled and the mes-
sages
Secure Mode Active
and
Remote
appear on the
front panel display.
Terminating
360B Track-
ing Mode
To terminate 360B VNA tracking receiver mode op-
eration, you must first return the 680XXB to local
control and then turn off the Secure mode.
Step 1
Turn off the 360B VNA. This returns the
680XXB to local control.
Step 2
On the 680XXB, press
SYSTEM
, then
Reset
. This turns off the Secure mode.
USE WITH
USE WITH A 360B VECTOR
OTHER INSTRUMENTS
NETWORK ANALYZER
680XXB OM
7-13/7-14