pad and appropriate termination key. When you
have finished setting the open parameter, close it by
pressing its menu soft-key or by making another
menu selection.
Press
Enable
to begin master-slave operation.
Press
CW
to return to the CW menu.
Master-Slave
Operation
During master-slave operation, the Slave unit is in
remote mode under the direct control of the Master
68XXXB. The Slave unit displays the following:
Its output CW frequency or sweep frequency
range.
Its output power level.
The messages
Remote
and
Local Lockout
.
The CW/sweep frequency settings on the Master
68XXXB define the master sweep, and the corre-
sponding frequency settings on the Slave unit
define the slave sweep. For example, if slave fre-
quency F1 is set to 4 GHz and slave frequency F2 is
set to 12 GHz, then the Slave unit will sweep from
4 to 12 GHz whenever the F1-F2 sweep range is se-
lected on the Master 68XXXB. The Master 68XXXB
will sweep from F1-F2 with the values of F1 and F2
defined in the Master units’s frequency list.
Master-Slave
Operation in
VNA Mode
In the VNA mode of master-slave operation, a Slave
unit is coupled to a Master 68XXXB that is con-
nected to a Model 360B Vector Network Analyzer in
a source or dual source configuration. (Operating in-
structions for the vector network analyzer can be
found in the Model 360B VNA Operation Manual,
P/N 10410-00110.) The following paragraphs de-
scribe how to set up both 68XXXBs to perform mas-
ter-slave operations in the VNA mode.
Place both 68XXXBs in CW mode. Then, on the Mas-
ter unit, press
Master Slave
to go to the Master-
Slave Menu display (page 7-5).
At the Master-Slave menu, press
Slave Freqs
to go
to the Slave Frequencies List Menu display (next
page).
USE WITH
MASTER-SLAVE
OTHER INSTRUMENTS
OPERATION
NOTE
The 562 SNA, when being used
with the master-slave configura-
tion, will not display markers.
680XXB OM
7-7
Summary of Contents for 680 B Series
Page 5: ......
Page 11: ...Figure 1 1 Series 680XXB Synthesized CW Generator...