Byte-No.
1
2
3
4
5
6
7
8
9
10
Format
Value
IEEE-754-
Flieskommazahl
PA-OUT
(FB1)
Status
Status
(FB1)
IEEE-754-
Flieskommazahl
Temperature
(FB2)
Status
Status
(FB2)
F
ig
.
19
:
T
elegram con
fi
guration example
2
E
xample
3
with pressure value and additional cyclical value without temperature value
:
l
AI
(
PA
-
OUT
)
l
F
ree
P
lace
l
A
dditional
C
yclic
V
alue
Byte-No.
1
2
3
4
5
6
7
8
9
10
Status
Status
Status
Status
(FB1)
IEEE-754-
Floating point value
PA-OUT
(FB1)
IEEE-754-
Floating point value
Additional Cyclic
Value
Format
Value
F
ig
.
20
:
T
elegram con
fi
guration example
3
D
ata format of the output signal
Byte4
Byte3
Byte2 Byte1 Byte0
Status
Value (IEEE-754)
F
ig
.
21
:
D
ata format of the output signal
T
he status byte corresponds to pro
fi
le
3
.
0
"
P
ro
fi
bus
PA P
ro
fi
le for
P
rocess
C
ontrol
D
evices
"
coded
.
T
he status
"
M
easured value
OK
"
is coded as
80
(
hex
) (
B
it
7
=
1
,
B
it
6
…
0
=
0
).
T
he measured value is transferred as a
32
bit
fl
oating point number in the
IEEE
-
754
format
.
Byte n
Byte n+1
Bit
7
VZ
Exponent
Bit
6
2
Bit
5
2
Bit
4
2
Bit
3
2
Bit
2
2
Bit
1
2
Bit
0
2
Bit
7
2
Bit
6
2
Bit
5
2
Bit
4
2
Bit
3
2
Bit
2
2
Bit
1
2
Bit
0
2
Sign
Bit
Significant
7
6
5
4
3
2
1
0
-1
-2
-3
-4
-5
-6
-7
Byte n+2
Byte n+3
Bit
7
2
Significant
Bit
6
2
Bit
5
2
Bit
4
2
Bit
3
2
Bit
2
2
Bit
1
2
Bit
0
2
Bit
7
2
Bit
6
2
Bit
5
2
Bit
4
2
Bit
3
2
Bit
2
2
Bit
1
2
Bit
0
2
Significant
-9
-10 -11 -12 -13 -14 -15 -16 -17
-18 -19
-20 -21 -22 -23
-8
Value = (-1)
VZ
2
(Exponent - 127)
(1 + Significant)
F
ig
.
22
:
D
ata format of the measured value
P
rocess pressure transmitter
IPT
-
11
V
ers
.
4
.
0
-
ceramic sensor
•
P
ro
fi
bus
PA
51
10
S
upplement
31547
-
EN
-
081211