background image

 

VI   RECOMMENDED DISINFECTING PROCEDURE  

(AFTER INSTALLATION IS COMPLETE) 

 

The storage tank lines going to the faucet must be disinfected to prevent bacteria contamination and 

possible odor problems from developing.  A dirty prefilter cartridge can also cause odors to develop.   

 

Trace amounts of hydrogen sulfide or methane gas dissolved in the raw water can also cause problems.  

(These gaseous compounds will permeate into and through the RO membranes). 

 

The method for proper disinfecting is by using household chlorine bleach and the P6UC4 tube fitting 

supplied with the RO unit in the accessory kit:  (Refer to Drawing.) 

 

1.  Unplug the RO unit.  The power must be off before disinfecting. 
2.  Turn the water supply off (cold water valve). 
3.  Drain the storage tank. 
4.  Disconnect the tube from the feedwater fitting going to the pump (step 1). 
5.  Disconnect the 1/4” tube on the check valve pressure tee switch from the unit (step 2). 
6.  Remove the 3/8” and 1/4” nuts, grab rings, spacers, and O-rings from the P6UC4 poly fitting.  

Put them aside and keep them as spares. 

7.  Connect the P6UC4 Poly fitting to the ends of the each of the tubing connections removed 

(step 3). The tubing connections are long enough to bend together but be careful not to kink 

any tubing connections.  Connecting these tubes will by-pass the pump during the disinfecting 

procedure. 

7.  Remove both of the carbon cartridges from the housings on the CAK-139 BP or the sediment 

filter from the housing on the NAK-139 BP.  Add 1 ounce of household chlorine bleach to the 

filter sump and place back on the unit.  Open the feedwater supply (cold water valve) and the 

RO faucet on the sink.  Run the water until a strong chlorine bleach smell is noted.  

8.  Close the RO faucet and fill the pressure tank.  Let the tank sit for 30 minutes, allowing the 

chlorine to kill any bacteria that might be present.  Drain the tank and close the cold-water 

valve. 

9.  Disconnect the P6UC4 Poly fitting and reconnect the tubing to their original connections. 

10.  Plug in the RO unit. 
11.  Open the cold-water valve and let the RO water to fill the pressure tank.  Drain and flush the 

tank several times to remove the chlorine.   DO NOT USE THE FIRST 2 TANK-FULLS OF 

WATER!! 

Summary of Contents for CAK Series

Page 1: ...C Ca ak k N Na ak k W Wa at te er r P Pu ur ri if fi ic ca at ti io on n S Sy ys st te em m REMOVE 90 TO 98 OF THE TOTAL DISSOLVED SOLIDS AND 100 OF PARTICULATE CONTAMINANTS...

Page 2: ...s manual be read and understood before performing installation operation and maintenance of the system The equipment warranty may be voided if instructions are not followed correctly The information p...

Page 3: ...s Pressurized Water Hazardous Chemicals Noise Failure to follow good safety practices and the instructions included in this manual may result in damage to the unit its surroundings personal injury or...

Page 4: ...ll out the operator s log sheet completely and return to XYLEM WET to validate warranty NOTE Hardness and iron will foul the membranes and result in a decrease in the amount and quality of the water p...

Page 5: ...ced Clogged inlet filters As pre filters age and collect debris the pressure drop across the filter increases and flow decreases At some point the reduced pressure or flow may actually cause the syste...

Page 6: ...tio Rejection of the TDS Total Dissolved Solids is at least 90 at 185 PSI 78 F and 1000 PPM TDS NOTE If the RO pump pressure exceeds 190 PSI 1 Remove the acorn nut on the right side of the pump 2 Usin...

Page 7: ...poly fitting Put them aside and keep them as spares 7 Connect the P6UC4 Poly fitting to the ends of the each of the tubing connections removed step 3 The tubing connections are long enough to bend to...

Page 8: ...y done with the sample port See the SYSTEM LAYOUT drawing NOTE Higher levels of chlorine may require a larger carbon prefilter if one is needed contact your local sales representative 3 RO PUMP once e...

Page 9: ...eat build up can also cause leaks to develop and will cause fouling of the membranes VENTILATION Inadequate ventilation will cause the pump motor to overheat and cut out It will also cause leaks due t...

Page 10: ...change in water production is approximately 2 Thus if the water temperature is 5 degrees C the amount of water produced will be 40 less than at 25 degrees C TDS AND REJECTION RATE As the TDS total di...

Page 11: ...drain the water from the RO membrane assembly and plug the holes before shipping NOTE 2 For the most accurate TDS readings let the RO unit run for at least 15 minutes NOTE 3 For the warranty to be va...

Page 12: ...12...

Page 13: ...13 Part Number 07 MS 1020 Rev 3 October 2018...

Reviews: