Bench-Top System Overview
Filter System
Filter Access Panel
Release Knobs
Used to tilt the air filter
system for easier access
Fine Filters
Used to filter air supply
Holes for optional
drain attachments
Access holes to attach
external drains to filter
water trap
Holes for optional drain attachment
Filter Access Panel
Release Knob (Front)
Solenoid
(non-user serviceable)
If the “Service” LED indicator is lit,
the two air filters should be replaced.
This condition occurs after 100,000
cleaning cycles.
See “Maintenance”
for filter change procedure.
To access filter assembly:
1. Power the unit off.
2. Remove the air supply hose.
3. Remove the side panel
4. Loosen the two release knobs and
slide the knobs down to tilt the
assembly outward from the
chassis.
5. Unscrew the water trap bowls and
replace each filter inside the
bowls.
NOTE: Periodically open the side
panel and check for condensation
inside the water traps. Open the
traps by loosening the drain knobs
on the bottom of the traps.
Filter Water
Traps