14
Bench-Top System Overview
System Status Indicators
Power/Ready
-
This LED will alert the user when there is AC power connected to the
system (RED LED), and it will indicate when the nozzle safety switch is released allowing
the cleaning cycle to be activated (GREEN LED). The cleaning cycle WILL NOT activate
when this LED is RED. This switch is similar to a safety release found on some power
tools, i.e. saws, nail guns, etc. To release the safety switch, apply slight pressure against
the bulkhead mating sleeve and gently slide the handset forward. Notice the POWER/
READY LED turn green. Press the RUN button on the handset.
Low Solvent
-
When lit, indicates the solvent level is too low to operate and the system
is in standby mode. The solvent resevoir must be refilled before operation can be contin-
ued. The solvent resevoir contains 225 ml. of solvent.
Service -
When lit, indicates the system’s air filters need to be replaced. The reset
button is pushed to reset the system’s cycle counter to zero. System will count 100,000
cleaning cycles before the filter change is necessary as indicated by this lit LED.
2. Press RUN button. It is not necessary to
hold the button down.
1. Apply slight pressure and
slide handset forward to
disable the safety switch.