Welch Allyn
使用说明
43
300, 250, 200, 150, 100, 50 (60 for US) and
0 mm Hg.
Note:
Your ability to measure the accuracy of a gauge
depends upon the sensitivity of the pressure standard
you use for the calibration procedure.
If using a manometer (mercury column or aneroid
gauge) rated at ± 3.0 mm Hg, an undetectable
error of up to 6.0 mm Hg is possible. If using a
device (e.g., digital pressure standard) rated at
± 0.1 mm Hg, an undetectable error of up to only
3.1 mm Hg is possible.
Welch Allyn recommends using as sensitive as
possible a pressure standard when performing
calibration checks. A Setra Pressure Meter (part
no. 2270-01), which is calibrated for ± 0.1 mm Hg,
or Netech (part no. 200-2000IN), which is
calibrated for ± 1.0 mm Hg, works well for this
application. Welch Allyn offers each meter, contact
distributor.
Warranty
Your Welch Allyn product, when new, is warranted
to be free from original defects in material and
workmanship and to perform in accordance with
manufacturer’s specifications under normal use
and service. The warranty period* begins from the
date of purchase from Welch Allyn, Inc. or its
authorized distributors. Welch Allyn’s obligation is
limited to the repair or replacement of components
determined by Welch Allyn to be defective within
the warranty period. These warranties extend to
the original purchaser and cannot be assigned or
transferred to any third party.
This warranty shall not apply to any damage or
product failure determined by Welch Allyn to have
been caused by misuse, accident (including
shipping damage), neglect, improper maintenance,
modification, or repair by someone other than
Welch Allyn or one of its authorized service
representatives.
*Gauge Warranty:
Should the Welch Allyn aneroid
sphygmomanometer deviate from the ± 3 mm Hg
accuracy specification during the warranty period,
Welch Allyn will recalibrate the
sphygmomanometer at no charge. Return the
gauge only to Welch Allyn Technical Service or
your local Welch Allyn authorized representative.
Welch Allyn Model 509:
Lifetime calibration warranty
Welch Allyn Model 767:
Lifetime calibration warranty
*Cuff Warranty:
Welch Allyn FlexiPort
TM
Reusable Cuff:
Three-year warranty.
Welch Allyn Two-Piece Cuff:
Two-year warranty.
These express warranties are in lieu of any and all
other warranties, express or implied, including the
warranties of merchantability and fitness for a
particular purpose, and no other person has been
authorized to assume for Welch Allyn any other
liability in connection with the sale of the Welch
Allyn product. Welch Allyn shall not be liable for
any loss or damages, whether direct, incidental, or
consequential, resulting from the breach of any
express warranty, except as set forth herein.
FlexiPort
TM
Connections
Model
509
Model
767
Summary of Contents for GS 777
Page 1: ...GS 777 GS 777 Integrated Wall System Directions for Use Material No 718848...
Page 2: ......
Page 6: ...vi Directions for Use Welch Allyn...
Page 8: ...2 Directions for Use Welch Allyn Welch Allyn 3 5V 1 2 3 2 10 4 3 5V 1 2 3 4 1 8 3 5 6 6...
Page 10: ...4 Directions for Use Welch Allyn...
Page 22: ...16 Directions for Use Welch Allyn...
Page 26: ...20 Directions for Use Welch Allyn...
Page 28: ...22 Directions for Use Welch Allyn...
Page 46: ...40 Directions for Use Welch Allyn...
Page 50: ...44 Directions for Use Welch Allyn...
Page 53: ......