Directions for Use

Welch Allyn

minimum of 30 seconds, repeat the above 
steps for a second reading.

Specifications

The Welch Allyn aneroid sphygmomanometer is 
accurate to ± 3 mm Hg.

This product will maintain the safety and 
performance characteristics specified at 
temperatures ranging from 10 °C to 40 °C at a 
relative humidity level not to exceed 90%.

Standards

• American National Standard ANSI/AAMI, SP10: 

2002, Manual, electronic, or automated 
sphygmomanometers.

• European Standard EN 1060-1: 1995, Non-

invasive sphygmomanometers-Part 1: General 
Requirements.

• European Standard EN 1060-2: 1996, Non-

invasive sphygmomanometers-Part 2: General 
Requirements.

• Meets the technical requirements of INMETRO 

Technical Metrological Regulation Number 24 of 
February 26, 1996.

Maintenance

Cleaning

Aneroid Gauge, Inflation Bulb, and Valve: Wipe the 
aneroid gauge, inflation bulb, and valve with 
slightly dampened cloth or alcohol pad.

Reusable One-Piece Blood Pressure Cuff:

 Use 

one or more of the following methods and allow to 
air dry:

• Wipe with mild detergent and water solution (1:9 

solution). Rinse.

• Wipe with Enzol per manufacturer’s instructions. 

Rinse.

• Wipe with 0.5% bleach and water solution. Rinse.

• Wipe with 70% isopropyl alcohol.

• Launder with mild detergent in warm water (60 °C 

max), normal wash cycle. Cuff is compatible with 
5 wash cycles (Reusable only). Close port cuff 
with laundering plug (part no. 5082-159).

Two-Piece Cuff and Bladder:

 Safely clean the 

cuffs with a damp cloth or wash in warm water (60° 
C maximum) with mild detergent. DO NOT PRESS 
WITH HOT IRON.

Before laundering the cuff:

1. Remove the bladder from the two-piece cuff.

2. Place the hook and loop fasteners in the 

closed position.

3. Machine launder using gentle cycle, warm 

water and mild detergent.

4. Air dry completely and reassemble 

components.

Low-level disinfection procedure (FlexiPort

TM

 

Reusable cuffs only):

Prepare Enzol® enzymatic detergent according to 
manufacturer's instructions. Apply port-cap 
(accessory part 5082-159) to cuff. Spray detergent 
solution liberally onto cuff and use a sterile brush to 
agitate the detergent solution over entire cuff 
surface for five minutes. Rinse continuously with 
distilled water for five minutes. To disinfect, first 
follow the cleaning steps above, then spray cuff 
with 10% bleach solution until saturated, agitate 
with a sterile brush over entire cuff surface for five 
minutes. Rinse continuously with distilled water for 
five minutes. Wipe off excess water with sterile 
cloth and allow cuff to air dry.

Calibration Check of Aneroid 
Sphygmomanometer

Quick Check of Calibration:

At zero pressure, make certain the pointer is within 
the oval surrounding the zero-pressure gradation 
on the dial.

Although an unpressurized reading of zero does 
not guarantee accuracy at all scale points, failure 
of the pointer to indicate zero (± 3 mm Hg) is an 
obvious sign of error.

Full Check of Calibration:

Welch Allyn recommends a full check of calibration 
at least every two years or according to local law 
and after maintenance and repair use the following 
procedure.

1. Connect the instrument under test to a high 

quality, known pressure standard and a 150 to 
500 cc test volume using a T-connector.

2. Pressurize gauge to slightly above 

 

300 mm Hg.

3. Bleed pressure down no faster than 10 mm Hg 

per second, stopping to check the pressure at 

Summary of Contents for GS 777

Page 1: ...GS 777 GS 777 Integrated Wall System Directions for Use Material No 718848...

Page 2: ......

Page 3: ...Welch Allyn iii Contents GS 777 1 1 1 1 1 2 2 3 3 3 3 3 SureTemp Plus 5 5 5 6 SureTemp Plus 7 7 7 8 8 8 8 8 9 9 10 10 10 10 11 11 11 11 12 12 12 12 12 13 13 13 13 13 13 13...

Page 4: ...Cleaning 25 Specifications 26 Environment 26 Additional Technical Information 26 Approvals 26 Warranty 26 SureTemp Plus 27 Symbols and Descriptions 27 Safety and Warnings 28 Introduction 29 SureTemp...

Page 5: ...mometer and Probe 37 Cleaning the Removable Probe Well 37 Disposal 38 Disposing of Equipment 38 Disposing of Batteries 38 Performance and Technical Specifications 38 Replacement Parts and Accessories...

Page 6: ...vi Directions for Use Welch Allyn...

Page 7: ...Welch Allyn 1 GS 777 Welch Allyn 3 5V Welch Allyn Rx only Welch Allyn Welch Allyn WEEE WEEE 2002 96 EC SJ T 11363 2006 BF 1 USB Welch Allyn 2 3 1 3 8 35 5 127...

Page 8: ...2 Directions for Use Welch Allyn Welch Allyn 3 5V 1 2 3 2 10 4 3 5V 1 2 3 4 1 8 3 5 6 6...

Page 9: ...llTransformer IEC 60601 1 UL 60601 1 IEC 60601 1 2 CAN CSA C22 2 No 601 1 M90 AS NZS3200 www welchallyn com GreenSeries777WallTransformer II 100 240V 50 60 Hz 0 18A 5V 1 A 5V 1 A 5V 0 5 A www welchall...

Page 10: ...4 Directions for Use Welch Allyn...

Page 11: ...Welch Allyn 5 SureTemp Plus REF 690 692 6 000 846 6 036 361 6 971 790 BF AA 1 5 MAX IPX Welch Allyn Welch Allyn Welch Allyn 3 5...

Page 12: ...6 Directions for Use Welch Allyn Welch Allyn SureTemp Plus Welch Allyn SureTemp Plus Welch Allyn Welch Allyn SureTemp Plus SureTemp Plus Welch Allyn Welch Allyn SureTemp Plus 690 692 SureTemp Plus...

Page 13: ...Plus SureTemp Plus 690 692 690 Welch Allyn SureTemp Plus Welch Allyn SureTemp Plus SureTemp Plus 690 SureTemp Plus 692 X X X X X X X X X X X X X X X X X X X X X X X 2 3 90 90 1 1 SureTemp Plus 690 Sur...

Page 14: ...8 Directions for Use Welch Allyn SureTemp Plus 1 5 AA 1 7 2 3 Welch Allyn 1 4 10 25 2 48 122 3 12 30 5 C F C F 692 0 15 30 45 60 60 5 Welch Allyn...

Page 15: ...Welch Allyn 9 SureTemp Plus 4 6 SureTemp Plus Welch Allyn 7 690 692 3 5 110 0 F 43 3 C 80 0 F 26 7 C 104 0 F 40 0 C 50 0 F 10 0 C...

Page 16: ...10 Directions for Use Welch Allyn 1 2 3 4 5 6 7 30 8 3 9 10 SureTemp Plus 10 13 17 12 15 18 1 2 3 Welch Allyn Welch Allyn 3 20 Welch Allyn...

Page 17: ...Welch Allyn 11 1 2 3 4 30 5 5 6 7 SureTemp Plus 10 13 1 2 3 4 1 5 5 8 5 6 30 Welch Allyn Welch Allyn 5 Welch Allyn Welch Allyn...

Page 18: ...12 Directions for Use Welch Allyn 7 3 8 9 10 3 5 1 2 3 4 5 6 7 8 9 10 SureTemp 692 25 50 100 200 SECURITY COUNT 5 4 3 2 1 10 SEC SECURITY COUNT OFF 3 Welch Allyn Welch Allyn 3 5 3 3 5 60...

Page 19: ...Welch Allyn 13 5 SEC 30 ON OFF OFF 1 2 3 4 3 AA 1 5 3 AA 1 5 5 6 70 10 CaviCide 70 10 CaviCide AA 1 5...

Page 20: ...3 690 2 Welch Allyn Inc Welch Allyn Welch Allyn 692 3 690 2 Welch Allyn Welch Allyn Welch Allyn Welch Allyn RMA Welch Allyn Welch Allyn Welch Allyn 1 2 Welch Allyn 3 4 5 50 0 F 104 0 F 10 0 C 40 0 C 5...

Page 21: ...Welch Allyn 15 1 Welch Allyn Welch Allyn 1 90 Welch Allyn Welch Allyn 90 WELCH ALLYN WELCH ALLYN...

Page 22: ...16 Directions for Use Welch Allyn...

Page 23: ...1 5 Welch Allyn 1 2 FlexiPortTM 3 4 5 1 Artery Index Marker Range 2 2 5 3 30 mm Hg 4 2 3 mm Hg 5 30 Welch Allyn 3 mm Hg 10 C 40 C 90 American National Standard ANSI AAMI SP10 2002 European Standard E...

Page 24: ...Hg 300 250 200 150 100 50 60 0 mm Hg 3 0 mm Hg 6 0 mm Hg 0 1 mm Hg 3 1 mm Hg Welch Allyn Setra 2270 01 0 1 mm Hg Netech 200 2000IN 1 0 mm Hg Welch Allyn Welch Allyn Welch Allyn Welch Allyn Inc Welch A...

Page 25: ...Welch Allyn 19 Welch Allyn Welch Allyn Welch Allyn Welch Allyn Welch Allyn Welch Allyn FlexiPortTM DuraShock Welch Allyn Inc FlexiPortTM 767 509...

Page 26: ...20 Directions for Use Welch Allyn...

Page 27: ...Welch Allyn 21 20000 Welch Allyn 20000 1 90 3 5 WelchAllyn 2 1 2 A 3 A 4 Welch Allyn Welch Allyn 3 1 2 11720 11730 11735 11470 11475 11710 3...

Page 28: ...22 Directions for Use Welch Allyn...

Page 29: ...gnetic interference and should not present problems to other equipment or be affected by other devices CAUTION Only authorized Welch Allyn Service Centers can perform service on this product CAUTION T...

Page 30: ...24 Directions for Use Welch Allyn Attach Welch Allyn 3 5 V instrument heads of your choice to the handles See the directions for use for each instrument head 2 3 1 3 8 35 5 127 4 1 8 3 5 6 6...

Page 31: ...utoclave the transformer or accessories WARNING Liquids can damage electronics inside the transformer Prevent liquids from spilling on or dripping into the transformer If liquids are spilled on or dri...

Page 32: ...d manufacturer s declaration information is available at www welchallyn com GreenSeries777WallTransformer Operating Temperature limitation Humidity limitation Atmospheric pressure limitation Transport...

Page 33: ...ed Alkaline Battery AA 1 5V IPX Not protected against the ingress of water MAX Latex Free Do not dispose of this product as unsorted municipal waste Prepare this product for reuse or separate collecti...

Page 34: ...ety and temperature measurement accuracy use only accessories and supplies recommended or supplied by Welch Allyn Equipment is not suitable for use in the presence of a flammable anesthetic mixture wi...

Page 35: ...y ingested medications age and gender A person s normal temperature tends to decrease with age The following table shows that ranges of normal also vary by site Do not directly compare measurements fr...

Page 36: ...mp Plus thermometer as shown below Wall Holder Standard X Pulse Timer X Backlight X Security Electronic X Configurable On Screen Labeling for Identification of Instrument X Sold in U S Canada X X Sold...

Page 37: ...gling of the probe cord with other equipment at the patient s bedside 1 Mount the instrument so that all other objects are no closer than 4 inches 10 25 cm from the sides and bottom of the unit 2 Moun...

Page 38: ...show that the Oral Mode is selected Adult Axillary Mode This flashing icon and or words are displayed to show that the Adult Axillary Mode is selected Ped Axillary Mode This flashing icon and or words...

Page 39: ...walking segments appear on the display indicating that measurement is in progress 7 The unit will beep three times when the final temperature is reached The measurement site temperature scale and pat...

Page 40: ...ccurate readings 3 Verify that the probe tip is completely surrounded by axillary tissue and place the arm snugly at the patient s side Hold the patient s arm in this position and do not allow movemen...

Page 41: ...separating the buttocks in place and hold the probe in place throughout the measurement process Rotating walking segments appear on the display indicating that measurement is in progress 6 The unit wi...

Page 42: ...osition after the temperature is displayed 6 Press the Mode Selection button until the Monitor Mode indicator appears on the LCD 7 Hold the thermometer in place for the required duration 8 Record the...

Page 43: ...ers from the probe cover storage compartment on the back of the unit 2 Remove the probe and the probe well 3 Press inward on the textured grips on each side of the assembly to remove the battery door...

Page 44: ...can perform or approve all repairs on products under warranty Unauthorized repairs will void the warranty Qualified electronics personnel or at a Welch Allyn Service Center should repair out of warran...

Page 45: ...risk Welch Allyn will not be responsible for loss associated with the use of any Welch Allyn product that 1 has had the serial number defaced 2 has been repaired by anyone other than an authorized Wel...

Page 46: ...40 Directions for Use Welch Allyn...

Page 47: ...in measurement error Connections 1 Connect one end of the coil tubing to the sphygmomanometer valve at the bottom of the unit 2 Align and press the FlexiPortTM adapter onto the cuff port 3 Attach the...

Page 48: ...ing the cuff 1 Remove the bladder from the two piece cuff 2 Place the hook and loop fasteners in the closed position 3 Machine launder using gentle cycle warm water and mild detergent 4 Air dry comple...

Page 49: ...warranties extend to the original purchaser and cannot be assigned or transferred to any third party This warranty shall not apply to any damage or product failure determined by Welch Allyn to have b...

Page 50: ...44 Directions for Use Welch Allyn...

Page 51: ...s Note It is important to note that while the throat illuminator section provides an excellent source of general illumination more then momentary direct skin contact should be avoided due to possible...

Page 52: ...46 Directions for Use Welch Allyn Figure 1 Figure 2 Ophthalmoscopes No 11720 11730 No 11735 Ophthalmoscopes No 11470 11475 No 11710 Align Align Figure 3...

Page 53: ......

Page 54: ...llyn Medical Equipment Suzhou Co Ltd Floor 3 Building 9 No 8 Jinfeng Road Suzhou Science Technology Town Jiangsu Province China 215163 Telephone 86 512 6808 6078 Fax 86 512 6808 7806 8 9 3 215163 86 5...

Reviews: