Using the Handles
Ensure that the transformer is plugged in.
1. Lift the handle from the cradle. The instrument
powers on. Only one instrument can be
powered at a time.
2. Adjust the light output by turning the rheostat
on the handle.
• Turning the rheostat clockwise increases
the light output.
• Turning the rheostat counterclockwise
decreases the light output.
Note:
Decreasing the light output to the
minimum level does not power down the
instrument.
3. Follow the directions for use for the instrument
head when examining a patient.
CAUTION:
To prevent damage, do not
overstretch the cords on the handles.
CAUTION:
To minimize the external housing
temperature of the instrument head, on-time
must not exceed 2 minutes, and off-time
must be a minimum of 10 minutes.
4. Return the handle to the cradle. The 3.5V
instrument powers down.
Cleaning
WARNING:
Before you clean the
transformer, unplug it from the wall socket.
WARNING:
Electric shock hazard. DO NOT
autoclave the transformer or accessories.
WARNING:
Liquids can damage electronics
inside the transformer. Prevent liquids from
spilling on or dripping into the transformer. If
liquids are spilled on or drip into the
transformer:
1. Disconnect the power plug.
2. Remove the transformer from the wall.
3. Dry off excess liquid from the transformer.
4. Mount the transformer on the wall.
5. Verify that the transformer functions normally
before using it.
Clean the transformer on a routine basis according
to your facility's protocols and standards or local
regulations.
To clean the transformer (excluding the instrument
heads), use any of the following:
• A clean cloth slightly dampened with 70 percent
isopropyl alcohol or 10 percent chlorine bleach
solution. Wipe dry with a clean cloth.
• Sani-Cloth
®
Plus.
• CaviWipes
™
(see Caution below).
CAUTION:
Do not use CaviWipes to clean
the blue handle cradles. This cleaning agent
produces bubbles and liquid during the
cleaning process that can enter openings in
the cradles and limit handle performance for
as long as 30 minutes after cleaning.
To clean the instrument heads, follow the
instructions provided in their directions for use.
Summary of Contents for GS 777
Page 1: ...GS 777 GS 777 Integrated Wall System Directions for Use Material No 718848...
Page 2: ......
Page 6: ...vi Directions for Use Welch Allyn...
Page 8: ...2 Directions for Use Welch Allyn Welch Allyn 3 5V 1 2 3 2 10 4 3 5V 1 2 3 4 1 8 3 5 6 6...
Page 10: ...4 Directions for Use Welch Allyn...
Page 22: ...16 Directions for Use Welch Allyn...
Page 26: ...20 Directions for Use Welch Allyn...
Page 28: ...22 Directions for Use Welch Allyn...
Page 46: ...40 Directions for Use Welch Allyn...
Page 50: ...44 Directions for Use Welch Allyn...
Page 53: ......