Weinmann Homecare MEDUCORE Easy Service And Repair Instructions Download Page 1

 

MEDUCORE

 

 

 

Easy

 

Automatic external defibrillator

 

Device nos. 1000 to 1340 from device no. 1341 with ILCOR 2005

 

Service and Repair instructions

 

MEDUCORE

 

 

 

Easy

 

 with Battery-Pack

 

MEDUCORE

 

 

 

Easy

 

 with Rechargeable Battery-Pack 

Summary of Contents for Homecare MEDUCORE Easy

Page 1: ...Easy Automatic external defibrillator Device nos 1000 to 1340 from device no 1341 with ILCOR 2005 Service and Repair instructions MEDUCORE Easy with Battery Pack MEDUCORE Easy with Rechargeable Batte...

Page 2: ...21 6 3 Testing devices 21 6 4 Replace SRC sticker 21 6 5 Documentation 21 7 Malfunctions and Rectification 22 8 Repair information and repair instructions 24 8 1 General 24 8 2 Tools and facilities 2...

Page 3: ...ying out repairs as shown in the service and repair manual In case of warranty claims all devices are to be sent to WEINMANN In order to handle warranty or goodwill requests we will require you to sub...

Page 4: ...trode plug 9 Carrying handle 10 Shock key Underside of device 11 Display Shock preperation 17 Battery Pack Rechargeable Battery Pack optional 16 Self sticking defibrillation electrodes 19 Emergency se...

Page 5: ...s for the delivery of an electric shock The delivery of a shock is carried out directly by the user after the device has prompted to do so Only use the device MEDUCORE Easy for the purposes described...

Page 6: ...activated metronome function gives off an acoustic metronome signal at a frequency of 100 beats per minute during the HLR pause Carry out the cardiac massage in rhythm with audible signal tone After...

Page 7: ...should carry out a heart lung revival Application documentation The MEDUCORE Easy saves ECG and event data The data could later be used e g in the follow up of treatment For this use the documentatio...

Page 8: ...his can otherwise lead to damage of the device and consequently endanger users and patients The MEDUCORE Easy can be kept properly clean by simply wiping to disinfect Please observe the users instruct...

Page 9: ...es of such errors and recommended countermeasures by referring to chapter 7 Malfunctions and Rectification on page 22 We recommend that you generally keep the following ready for use Emergency Set WM...

Page 10: ...ings cannot be modified 4 3 Entering the device data Enter the serial number of the device Device No and the manufacturing date into the test records 4 4 Checking the devices self testing 1 Insert rec...

Page 11: ...input shock button and shock output Preparation of the MEDUCORE Easy 1 Connect the rechargeable battery adapter cur rent measuring device and rechargeable battery or battery pack as illustrated 2 Inst...

Page 12: ...e Press the flashing shock key activate this Requirement The shock is triggered by activation of the shock button The shock is given The device clearly gives the spoken message Shock was delivered The...

Page 13: ...illator tester Sinus EKG Request All of the LEDs in the green pentagon light up 4 8 Checking the info button and reading contact 1 While the green pentagon is lit up press the info button Request The...

Page 14: ...is correctly attached easy to open and closes firmly Functions open close switch on are ok Inner part The socket for the electrode plug is free from dirt Rechargeable Battery Pack Completely charged...

Page 15: ...ance and SRC stickers are affixed correctly SRC sticker applies only to Germany If repairs have been carried out then you should replace the stickers with new ones 4 13 Preparing a device for dispatch...

Page 16: ...Set the date and time with the help of the EasyView see chapter Operation in the operat ing instructions for the PC Software EasyView WM 16880 After exchanging the printed circuit board if no customer...

Page 17: ...Checking the device 17 4 14 Documentation Note down the points 4 3 to 4 12 as well as the test date and tester number in the test records...

Page 18: ...the Battery Rechargeable Battery see chapter 5 2 page 18 Replace internal battery see chapter 8 8 page 29 Final check in accordance with test directive WM 40006 see chapter 4 page 9 Renew maintenance...

Page 19: ...checks the charging level Offer your customer a new Battery Pack Recommend that your customer replaces the Battery Pack as soon as possible Capacity is 8 Inform your customer about the charging level...

Page 20: ...ase contact a licensed and certified electronic waste recycler Names and addresses can be obtained from your Environmental Officer or municipal authorities Disposal of batteries Do not dispose of spen...

Page 21: ...an additional description of the required steps to be taken 6 3 Testing devices Additional to the named testing devices described in chapter 4 1 you will need SRC protocol see WM 40007f page 2 on page...

Page 22: ...low Battery power levels are low Finish off current treatment and then replace batteries see Chapter Energy Supply in the operating instructions MEDUCORE Easy The red LED will light up flashing when...

Page 23: ...ce and check connections Chapter 8 3 Page 25 Main circuit board defective Open up device and exchange the main circuit board Chapter 8 9 Page 30 Shocks can not be delivered in spite of flashing shock...

Page 24: ...ing repairs After every repair a maintenance check must be performed see 4 Checking the device on page 9 When replacing components or single parts use only genuine WEINMANN Parts When ordering the dev...

Page 25: ...MEDUCORE Easy 2 Lie the MEDUCORE Easy with the red lid facing down on a non slip surface 3 Remove the sealing plugs 33 4 Loosen and remove the ten screws 30 5 Take the MEDUCORE Easy in both hands and...

Page 26: ...he front of the device The capacitor is very slightly charged The display will now show for example 0 07 Upon releasing the button the display must slowly return to 0 00 Danger Risk of electric shock...

Page 27: ...5 Pull the loudspeaker cable from the printed circuit board 6 Pull the red cable plug out of the printed circuit board 7 Pull the capacitor out of the lower encasing part and place it carefully on th...

Page 28: ...circuit board 5 Press the capacitor into the compartment of the lower encasing part Guide the cable in through the slit 6 Plug the cables for capacitor 27 onto the board Ensure that polarity is corre...

Page 29: ...he printed circuit board Make sure that the polarity is correct see the labelling on the printed circuit board and on the capacitor If the cables are mixed up then the upper and lower device parts wil...

Page 30: ...the capacitor onto the printed circuit board Make sure that insulation mats 28 are lying between the capacitor and the main circuit board 3 Plug the capacitor cable 27 into the printed circuit board M...

Page 31: ...he device configuration The number will now be stored in MEDUCORE Easy If a 4 digit serial number is entered the system will automatically pad it out to 5 digits by adding a leading 0 If the serial nu...

Page 32: ...circuit board 4 Plug the new cable in its place 5 Place the cable lug and spring washer onto the screw and screw the nut onto it Make sure that the order is correct 6 Close the device see Chapter 8 6...

Page 33: ...rged see Chapter 8 3 Page 25 1 Discharge capacitor and open up the device see Chapter 8 3 Page 25 2 Remove the main circuit board see Chapter 8 9 Page 30 3 Install the main circuit board into a new up...

Page 34: ...07 WM 40517 WM 40527 WM 40667 WM 40677 WM 40687 WM 40558 16 Defibrillation electrodes DE GB Defibrillation electrodes FR Defibrillation electrodes IT Defibrillation electrodes TH Defibrillation electr...

Page 35: ...rd RU new Main circuit board RU replacement Main circuit board PL new Main circuit board PL replacement Main circuit board NL new Main circuit board NL replacement Main circuit board ES new Main circu...

Page 36: ...M 40167 33 Sealing plug WM 40158 34 Battery 3 V WM 40089 35 Handle WM 40103 36 Handle sealing WM 40186 37 Cylinder head screw M8x30 DIN EN ISO 4762 ST ZN WM 50607 38 Flexible handle including attachin...

Page 37: ...MEDUCORE Easy JA TH Instructions for use MEDUCORE Easy PL RU Instructions for use MEDUCORE Easy IS FI Instructions for use MEDUCORE Easy TR GR Instructions for use MEDUCORE Easy ID CN Instructions for...

Page 38: ...Battery Adapter WM 40008 Test line MEDUCORE Easy WM 40454 10 3 Software EasyView Service PC software 10 4 Testing devices Multimeter measuring range up to 20 A Defi tester ECG graphs Sinusoidal rhyth...

Page 39: ...ss IEC 529 IPX4 protected against sprayed water Vibration and knock DIN EN 1789 1999 Free fall EN 60601 1 1996 Electromagnetic compatibility EN 60601 1 2 2001 EN 55011 1998 A1 EN 55014 1 2000 A1 EN 61...

Page 40: ...Stand by Time up to 5 years 4 years Maximum charging time 4 h Maximum charging power 1 2 A Transport Storage Temperature range Air humidity 40 C to 85 C 0 to 90 Service life 300 full charges Battery...

Page 41: ...ay affect or prevent the proper identification of arrhythmias It is thus possible that not all defibrillatable rhythms will be identified and shock output from the device is not recommended under cert...

Page 42: ...high Energy protocol constant rising Metronome function on off Data management Recommended safe distance between mobile HF telecommunications devices e g mobile telephones and the MEDUCORE Easy Nomin...

Page 43: ...e V1 2 2 WM 40192a 08 12 2006 Adjusting the EasyView software for the extended flash memory erase time on the WM 40130 PCB revision by the manufacturer V 1 2 3 WM 40192b 21 02 2007 Adjusting the EasyV...

Page 44: ...___________ __________________ Date Signature Device type MEDUCORE Easy Order No ________________________ Company _____________ __________________ Date Signature Date of manufacture ________________ S...

Page 45: ...Manufacturer s designation Operator Safety related check Due date 2 years Scope Verify that the equipment is complete Visual inspection for mechanical damage Test of system components Safety related c...

Page 46: ...y and capacitor discharge LEDs belonging to the progress displays pentagon triangle charge and shock button are all lighting correctly The capacitor is discharged if the shock is not delivered 8 Check...

Page 47: ......

Page 48: ...weg 40 D 22525 Hamburg T 49 0 40 5 47 02 0 F 49 0 40 5 47 02 461 E info weinmann de www weinmann de Center for Production Logistics Service Weinmann Ger te f r Medizin GmbH Co KG Siebenst cken 14 D 24...