Brine Nebulization System, Type Soldos V3
Index: 03
Change date: 01/03/2017
Operating instructions no.: BA DW 005-03 Soldos V3 EN.docx
Page 52 of 70
Test no. 1 to 5
The word Test appears in a green font in the field and the timer counts from 30
seconds to 0.
An electrical signal is provided at the output during this time. A consumer
connected to this output will operate during these 30 seconds.
Press the Button Test to cancel the test procedure.
Test no. 6 to 10
6.6.3
Pressure sensor test menu
This test is used to check the function of the pressure sensor.
Press start for Pressure Sensor 1 or Pressure Sensor 2 (version with 2 dosing
pumps).