3.5 Electrical Service
Figure 9
–
Location of power and interlocks
Figure 10
– Power and interlocks
The Element requires a 120VAC single phase electrical connection going into the rear
power and interlock panel. The “neutral” connection from the building power supply should
be connected to the upper 30A circuit breaker, and the “hot/line” connection should be
connected to the lower 30A circuit breaker. Perform electrical connections according to all
applicable electrical code
s. Ensure a suitable “Earth” connection is securely installed to
the designated green terminal block above/next to the 30A circuit breakers.
NOTE:
The circuit breakers located on the back of Element are not substitutes for a
building’s circuit overload protection! Ensure that properly rated conductors and matching
circuit breakers are used to supply the Element’s power and interlock panel.