Note
: Bytes 1-6 may not be visible, depending on the Master system used.
Channel related diagnostic data
Byte 1
Header
Byte 2
Channel type
Byte 3
Channel related error code
Channel related diagnostic data
Byte 3
Global error
=0xA9 (General error)
Over current
=0xA1 (Short circuit)
Under voltage
=0xA2 (Under voltage)
Over voltage
=0xA3 (Over voltage)
Over temp
=0xA5 (Over temp)
Motor stall
=0xA4 (Overload)
Tacho fault
=0xB1 (Device related 0x11)
Leak detected
=0xB2 (Device related 0x12)
Setpoint out of range - low
=0xA8 (Lower limit exceeded)
Setpoint out of range - high
=0xA7 (Upper limit exceeded)
Fluid level alert
=0xB3 (Device related 0x15)
Note: Global error format: 8 bit header, 8 bit channel type and 8 bit error type
Operating this pump under PROFIBUS control is beyond the scope of this instruction manual. Consult
your PROFIBUS network literature for further information.
Standard - 9 way D - PROFIBUS use
Interfacing the pump with the PROFIBUS network is by means of a 9 way D connector on the rear of the
pump, (within the N module if the pump is N variant). Cable suitable for use in a PROFIBUS DP
installation must be used.