Figure 3–13: Inlet manifold outlet-tubing fitting
1
2
3
Inlet manifold outlet-tubing fitting
Inlet manifold
i
2
Valve
12. Power-on the solvent manager.
13. Prime the solvent manager (see
Priming the quaternary solvent manager
3.11 Replacing the low-pressure inlet filters on the GPV
outlet
Warning:
Observe Good Laboratory Practice (GLP) at all times, particularly when
working with hazardous materials. Consult the Material Safety Data Sheets regarding
the solvents you use. Additionally, consult the safety representative for your
organization regarding its protocols for handling such materials.
Warning:
To avoid personal contamination from contact with biologically hazardous or toxic
materials, do not attempt to force or draw liquid or gas through the valve's inlet or outlet ports.
Notice:
To avoid contaminating system components, wear clean, chemical-resistant,
powder-free gloves when performing this procedure.
Requirement:
Wear clean, chemical-resistant, powder-free gloves when performing
this procedure.
Required tools and materials
• Chemical-resistant, powder-free gloves
• Low-pressure inlet filters (4)