B-10 External Connections
Sample manager I/O signal connectors
The rear panel of the sample manager includes a removable connector that
holds the screw terminals for I/O signal cables. This connector is keyed so that
it can receive a signal cable inserted only one way.
Requirement:
A contact closure output connection (Inject Start Out) from the
sample manager is required to trigger a mass spectrometer or an ACQUITY
2996 PDA detector running under MassLynx software control to start.
Sample manager I/O signal connectors:
For electrical specifications, see the
ACQUITY UPLC I-Class System
Specifications
document.
Chart-out signal conditions
Signal
Parameter Setting at 0
Volts (Minimum)
Parameter Setting at
2.000 Volts (Maximum)
Flow Rate
0.000 mL/min
2 mL/min
System Pressure
–345 kPa (–3.45 bar,
–50 psi)
124,106 kPa (1241 bar,
18,000 psi)
Composition
0.0%
100.0%
Injec
t S
tart
Out +
Injec
t S
tart
Out -
Gro
und
Gro
und
Inje
ct Hold
In +
Inje
ct Hold
In -
1 2 3 4 5 6