ACQUITY control panels
3-7
You can access these additional functions by right-clicking anywhere in the
binary solvent manager control panel:
Status
Displays the status of the current operation.
System Pressure
Displays system pressure, in kPa, bar, or psi.
You can customize pressure units via the
ACQUITY UPLC Console.
Flow Rate
Displays the total flow rate of the binary
solvent manager, from 0.000 to 2.000 mL/min
under normal operation and 0.000 to 8.000
mL/min when priming.
Solvent Composition
Displays the percentage of solvent to be
drawn from the solvent lines (A and B).
Composition values range from 0.0 to
100.0%.
(Stop Flow)
Immediately stops all flow from the binary
solvent manager.
Additional functions in the binary solvent manager control panel:
Control panel function
Description
Start up system
Brings the system to operational conditions
after an extended idle period or when
switching to different solvents.
See
“Starting up the system” in the
Binary
Solvent Manager Operator’s Overview and
Maintenance Information
.
Prime solvents
Displays the Prime Solvents dialog box.
See
“Priming the binary solvent manager” in
the
Binary Solvent Manager Operator’s
Overview and Maintenance Information
.
Prime seal wash
Starts priming the seal wash.
See
“Priming the seal wash system” in the
Binary Solvent Manager Operator’s Overview
and Maintenance Information
.
Binary solvent manager control panel items: (Continued)
Control panel item
Description