3-8
Preparing the System
Sample manager control panel
The sample manager control panel displays current sample compartment and
column heater temperatures and set points. You can edit these values when
the system is idle by clicking the underlined value. You cannot edit sample
manager set points while the system is running samples.
Tips:
•
To keep the sample compartment at a constant temperature, open its
door only when necessary.
•
The sample manager’s fans stop circulating air whenever the sample
compartment door is open.
Wash plungers
Initiates the plunger wash sequence, which
fills and then slowly empties the primary and
accumulator chambers (with the current
solvent composition) while performing a high
speed/volume seal wash. This action helps
prevent the buildup of precipitates on the
pump plungers which can damage the high
pressure seals.
Launch ACQUITY UPLC
Console
Launches the ACQUITY UPLC Console.
Reset BSM
Resets the binary solvent manager after an
error condition.
Help
Displays the ACQUITY UPLC Console
online Help.
Additional functions in the binary solvent manager control panel:
Control panel function
Description