December 5, 2017, 715004753 Rev. B
Page 58
4.2.4
Detector troubleshooting
Table 4–3:
2998 PDA detector troubleshooting
Symptom
Possible cause
Corrective action
Both LEDs unlit
No electrical power
1. Inspect line cord
connections.
2. Test electrical outlet for
power.
Open (spent) or defective fuse Replace fuse (see
Change in reference spectrum Mobile phase contains gas or
is contaminated
Prepare fresh mobile phase,
and degas thoroughly.
Air bubbles trapped in flow cell Reseat and check alignment
of flow cell.
Flush the flow cell (see
), or apply slight
backpressure of 207 to
345 kPa (2 to 3 bar, 30 to
50 psi) on the detector waste
outlet. For example, connect a
1- to 2-foot (30- to 60-cm)
length of 0.009-inch (0.23-mm)
ID tubing connected to the
detector waste outlet.
Continuous audible tone
Detector failure
Cycle power to the detector.
Detector not responding to
console
Bad or disconnected cable
Inspect cable connections,
tighten connectors, or replace
cable.
Configuration problem
Check Ethernet configuration.
For details, see Empower
online Help.
Lamp light flashes green and
power light constant green
The detector is initializing.
No corrective action required.
Wait until initialization
completes (both lights
constant green).
Summary of Contents for 2998
Page 2: ...December 5 2017 715004753 Rev B Page ii...
Page 8: ...December 5 2017 715004753 Rev B Page viii...
Page 14: ...December 5 2017 715004753 Rev B Page xiv...
Page 26: ...December 5 2017 715004753 Rev B Page 26...
Page 44: ...December 5 2017 715004753 Rev B Page 44...
Page 60: ...December 5 2017 715004753 Rev B Page 60...