background image

December 5, 2017, 715004753 Rev. B

Page 27

2

Setting Up the Detector

The chapter provides the information necessary to set up the 2998 PDA detector.

2.1

Before you begin

Requirement:

 To install the 2998 PDA detector, you must know how, in general, to set up and 

operate laboratory instruments and computer-controlled devices and also how to handle solvents.

Before installing the detector, ensure that

the required components are present;

none of the shipping containers or unpacked items are damaged.

2.2

Unpacking and inspecting

The carton in which the detector is shipper contains these items:

Certificate of structural integrity

2998 PDA detector

2998 Photodiode Array Detector Overview and Maintenance Guide 

(this document)

Startup kit

Release notes

When unpacking shipping cartons, ensure that their contents appear listed on the packing slips 
and that, conversely, the items listed on the packing slips are among the contents.

Recommendation: 

Save the shipping cartons for use in future transport or shipment. 

If you discover any damage or discrepancy when you inspect the contents of the cartons, 
immediately contact the shipping agent and your local Waters representative.

Customers in the USA and Canada should report damage and discrepancies to Waters Technical 
Service (800 252-4752). Others should phone their local Waters subsidiary or Waters corporate 
headquarters in Milford, Massachusetts (USA), or visit http://www.waters.com and click Offices.

For complete information on reporting shipping damages and submitting claims, see 

Waters 

Licenses, Warranties, and Support Services

.

Warning: 

To avoid injury, Waters recommends that two people lift the 2998 PDA detector.

Summary of Contents for 2998

Page 1: ...2998 Photodiode Array Detector Overview and Maintenance Guide 715004753 Revision B Copyright Waters Corporation 2015 2017 All rights reserved...

Page 2: ...December 5 2017 715004753 Rev B Page ii...

Page 3: ...or the most recent revision of this document consult the Waters website waters com Trademarks Waters Waters Quality Parts THE SCIENCE OF WHAT S POSSIBLE ACQUITY ACQUITY Arc Alliance Empower and MassLy...

Page 4: ...n all cases where the symbol is used to find out the nature of the potential hazard and any actions which have to be taken Considerations specific to the 2998 Photodiode Array Detector Power cord repl...

Page 5: ...strument so that it is difficult to disconnect the power cord Equipment misuse notice If equipment is used in a manner not specified by its manufacturer protections against personal injury inherent in...

Page 6: ...and above normal levels of a compound If sample trays are the same or very similar vary the location of the QC samples in the trays Ensure that QC sample results fall within an acceptable range and e...

Page 7: ...dustrial Scientific and Medical ISM instrument requirements Group 1 products apply to intentionally generated and or used conductively coupled radio frequency energy that is necessary for the internal...

Page 8: ...December 5 2017 715004753 Rev B Page viii...

Page 9: ...i Intended use of the 2998 Photodiode Array Detector vi Calibrating vi Quality control vi EMC considerations vii Canada spectrum management emissions notice vii ISM Classification ISM Group 1 Class B...

Page 10: ...ices 34 2 7 Plumbing the 2998 PDA detector 38 2 7 1 Connecting the detector to the gas supply 39 2 8 Starting up and shutting down the 2998 PDA detector 40 2 8 1 Starting the detector 40 2 8 2 Monitor...

Page 11: ...hooting 58 5 Spectral Contrast Theory 61 5 1 Comparing absorbance spectra 61 5 2 Representing spectra as vectors 62 5 2 1 Vectors derived from two wavelengths 62 5 2 2 Vectors derived from multiple wa...

Page 12: ...tion 83 C 1 1 Clean solvents 83 C 1 2 Solvent quality 83 C 1 3 Preparation checklist 83 C 1 4 Water 83 C 1 5 Use buffers 84 C 1 6 Tetrahydrofuran 84 C 2 Solvent miscibility 84 C 2 1 How to use miscibi...

Page 13: ...December 5 2017 715004753 Rev B Page xiii C 9 3 Wavelength selection for chromophore detection 91 C 9 4 Mobile phase absorbance 92...

Page 14: ...December 5 2017 715004753 Rev B Page xiv...

Page 15: ...le light UV Vis spectrophotometer With a photodiode array of 512 photodiodes and an optical resolution of 1 2 nm pixel the detector operates within a range of 190 to 800 nm The figure below illustrate...

Page 16: ...a deviation from absorbance versus concentration linearity The absorbance of mobile phase can reduce the linear range Table 1 1 Optics assembly components Component Function Lamp Deuterium source lamp...

Page 17: ...e entrance of the flow cell and a taper to the internal bore of the flow cell prevents light rays from striking the internal walls of the flow cell The Waters TaperSlit flow cell so called because of...

Page 18: ...n the array The following figure shows an absorbance spectrum of benzene Note that the wavelength resolution is sufficient to resolve five principal absorption peaks Figure 1 4 Benzene spectrum at dif...

Page 19: ...electing the appropriate sampling rate A sufficient number of points must fall across a peak to define its shape Thus the definition between peaks is lost at very low sampling rates Empower uses the e...

Page 20: ...the filter time constant that you select Increasing the filter time constant reduces baseline noise improving the signal to noise ratio Nevertheless increasing the constant too much artificially broad...

Page 21: ...e effects Greatly decrease baseline noise Shorten and broaden peaks For the highest resolution select an appropriate sampling rate and choose a fast filter time constant For the highest sensitivity se...

Page 22: ...remains the same 1 5 Computing absorbance data points The absorbance data points are calculated by the detector and then transmitted to the database MassLynx or Empower software The absorbance calcula...

Page 23: ...t as the absorbance data 1 5 3 Data averaging The data the detector reports to the database Empower or MassLynx can be the average of data points After calculating absorbance the detector averages abs...

Page 24: ...ingle data point The baseline noise decreases as the number of exposures sample increases Tip The amount of data stored is based on wavelength range spectral resolution runtime and sample rate Specify...

Page 25: ...s that you select It then subtracts that value from the absorbance value Abs Comp t Abs t CRef t where Abs Comp absorbance compensated Abs absorbance CRef compensation reference t elapsed time after s...

Page 26: ...December 5 2017 715004753 Rev B Page 26...

Page 27: ...kit Release notes When unpacking shipping cartons ensure that their contents appear listed on the packing slips and that conversely the items listed on the packing slips are among the contents Recomme...

Page 28: ...e to allow proper function of the drip management system drain tube to which you can connect a waste reservoir that diverts solvent leaks from the flow cell 2 4 Stacking system modules This procedure...

Page 29: ...and connected near the system To connect to the electricity source Recommendation When connecting to the electricity source use a line conditioner or an uninterruptible power supply UPS for optimum l...

Page 30: ...teady green when the lamp is ignited 2 6 Connecting signal cables See also Waters Ethernet Instrument Getting Started Guide The following figure shows the rear panel location of the connectors used to...

Page 31: ...detector Chart mark Adds a chart mark at 10 of full scale to either or both analog output channels Signal Out 1 and Signal Out 2 and is configurable Auto zero Calculates an offset value that when add...

Page 32: ...that accompany the instrument control software 2 6 1 1 Single Waters instrument connection In a single Waters instrument system configuration the connection hardware requires only one standard Etherne...

Page 33: ...multiple Ethernet instruments Network hubs are not supported Figure 2 6 Mixed system of IEEE and Ethernet instruments An inject start trigger is needed for the 2998 PDA detector Figure 2 7 All Etherne...

Page 34: ...ignal connections between the 2998 PDA detector s rear panel and these devices Waters Alliance Separations Module Waters eSAT IN Module Waters or other manual injector Other manufacturers integrator o...

Page 35: ...he generation of a stop flow connect cables shown in the following table and figure Requirement To automatically stop the solvent flow within the system in the event of an error condition or hardware...

Page 36: ...separations module Alliance separations module connector B 2998 PDA detector connector B Pin 4 stop flow red Pin 6 switch 1 red Pin 5 stop flow black Pin 7 switch 1 black Table 2 6 2998 PDA detector...

Page 37: ...m an injector it performs the corresponding auto zero or inject start function To send an auto zero or chart mark signal from an injector to the detector connect cables as shown in the following table...

Page 38: ...rs plastic or cloth covered Compression screw assemblies 3 To make plumbing connections to the detector 1 Measure the length of tubing needed to connect the column outlet to the detector inlet Warning...

Page 39: ...a quarter turn 6 Seat the free end of the column outlet tubing in the fitting of the detector inlet and then tighten the reusable finger tight fitting as in step 5 7 Seat the end of the detector outle...

Page 40: ...cause the detector to fail the reference energy diagnostic test Ensure that the solvent is correctly composed and that it is of high quality and miscible with any other solvents used in the system Use...

Page 41: ...10 mL of degassed methanol if the mobile phase does not contain buffers 2 Power off the detector 2 9 Using a cuvette The 2998 PDA detector cuvette option allows for ease of use in Sample handling Warn...

Page 42: ...using the cuvette when the HPLC instruments are in the idle or static state and operating under identical flow cell conditions 2 9 1 Before you begin Recommendation To ensure accurate results for the...

Page 43: ...the slot and does not slide forward Ensure that the cuvette does not become dislodged when you are replacing the cuvette holder 4 Gently guide the cuvette holder back into the flow cell assembly until...

Page 44: ...December 5 2017 715004753 Rev B Page 44...

Page 45: ...Offices When you contact Waters be prepared to provide this information Nature of the symptom Instrument serial number Solvent s Method parameters sensitivity and wavelength Type and serial number of...

Page 46: ...lockages in the solvent lines and flow cell Damage to the detector components Microbial growth 3 4 Maintaining the flow cell The flow cell of the 2998 PDA detector requires maintenance when these even...

Page 47: ...ffers flush the cell with 10 mL of HPLC grade water and then flush with 10 mL of a low surface tension solvent such as methanol Requirement Ensure that the solvent is miscible with the previous mobile...

Page 48: ...ly s front plate Grasp the handle and gently pull the flow cell assembly toward you 7 Unpack and inspect the new flow cell 8 Square the flow cell assembly in front of the opening and then insert it in...

Page 49: ...sible and then tighten slightly more using the 1 4 inch flat blade screwdriver 12 Connect the column outlet tubing to the flow cell inlet 13 Connect the waste line to the flow cell outlet 14 Close the...

Page 50: ...remely hot during operation Warning To avoid eye injury from ultraviolet radiation exposure power off the detector before changing the lamp wear eye protection that filters ultraviolet light keep the...

Page 51: ...amp dialogue box type the serial number of the new lamp see the label attached to the lamp connector wire and then click OK 3 6 Replacing the fuses The 2998 PDA detector requires two 100 to 240 Vac 50...

Page 52: ...s of spring loaded fuse holder 3 With minimum pressure withdraw the spring loaded fuse holder Figure 3 6 Removing fuse holder 4 Remove and discard the fuses 5 Ensure that the new fuses are properly ra...

Page 53: ...cell Before you power on the detector prime the system to fill the flow cell with solvent and remove any air To ensure that the detector is properly aligned and calibrated fill the flow cell with sol...

Page 54: ...calibration You can also start this procedure manually Notes This procedure can adversely affect spectral library matching and peak purity calibrations You might need to reacquire spectral libraries...

Page 55: ...oubleshoot Rear panel Result The PDA Detector Rear Panel dialog box appears The following table describes the 2998 PDA detector I O connections Table 4 1 2998 PDA detector analog out event in connecti...

Page 56: ...correct If after running the diagnostic functions applicable to your problem and troubleshooting the detector you cannot correct an error condition contact Waters Technical Service department 4 2 1 Tr...

Page 57: ...analysis Table 4 2 Lamp LED indications LED mode and color Description Unlit Indicates that the detector lamp is extinguished Constant green Indicates that the detector lamp is ignited and that the de...

Page 58: ...ush the flow cell see page 47 or apply slight backpressure of 207 to 345 kPa 2 to 3 bar 30 to 50 psi on the detector waste outlet For example connect a 1 to 2 foot 30 to 60 cm length of 0 009 inch 0 2...

Page 59: ...2 to 3 bar 30 to 50 psi on the detector waste outlet For example connect a 1 to 2 foot 30 to 60 cm length of 0 009 inch 0 23 mm ID tubing connected to the detector waste outlet Weak lamp Replace the...

Page 60: ...December 5 2017 715004753 Rev B Page 60...

Page 61: ...When measured at specific solvent and pH conditions the shape of a compound s absorbance spectrum characterizes the compound The varying extent of UV Vis absorbance occurring at different wavelengths...

Page 62: ...ce spectral sensitivity set the bench resolution to 1 2 nm Tip Do not include wavelengths where there is no analyte absorbance 5 2 1 Vectors derived from two wavelengths The spectral contrast algorith...

Page 63: ...versely if any comparison of ratios does not match the corresponding vectors point in different directions 5 3 Spectral contrast angles Spectra of identical shape have vectors that point in the same d...

Page 64: ...figure the absorbance spectra of two compounds A and B are similar They therefore produce a small spectral contrast angle 3 0 Figure 5 4 Spectra with a small spectral contrast angle Compound A Compou...

Page 65: ...contrast angle between these absorbance spectra of the same compound is 3 4 Figure 5 5 Normalized absorbance spectra of a compound at two concentrations 5 4 Undesirable effects Shape differences betw...

Page 66: ...tometric error for all spectral contrast operations the maximum spectral absorbance of a compound should be less than 1 AU Note that the absorbance of the mobile phase reduces the working linear dynam...

Page 67: ...can help determine whether the shape difference between spectra is significant In general a spectral contrast angle less than its threshold angle indicates that shape differences are attributable to n...

Page 68: ...December 5 2017 715004753 Rev B Page 68...

Page 69: ...the failure of individuals to comply with any safety precaution when installing repairing or operating any of its instruments or devices The following symbols warn of risks that can arise when you ope...

Page 70: ...se precautions when working in the vicinity of such tubing when it is pressurized Wear eye protection Extinguish all nearby flames Do not use tubing that is or has been stressed or kinked Do not expos...

Page 71: ...ures Warning To avoid personal contamination with biohazards toxic materials or corrosive materials you must understand the hazards associated with their handling Guidelines prescribing the proper use...

Page 72: ...ktionst chtigkeit verantwortlichen Personen kann zum Entzug der Bedienungsbefugnis des Systems f hren Avvertenza qualsiasi modifica o alterazione apportata a questa unit e non espressamente autorizzat...

Page 73: ...d k nnen nichtmetallische Schl uche quellen dadurch wird der Berstdruck des Schlauches erheblich reduziert Attenzione fare attenzione quando si utilizzano tubi in materiale polimerico sotto pressione...

Page 74: ...ion assur e par le mat riel risque d tre d fectueuses Vorsicht Der Benutzer wird darauf aufmerksam gemacht dass bei unsachgem er Verwendung des Ger tes die eingebauten Sicherheitseinrichtungen unter U...

Page 75: ...ez toujours les fusibles par d autres du type et de la puissance indiqu s sur le panneau proximit du couvercle de la boite fusible de l instrument Vorsicht Zum Schutz gegen Feuer die Sicherungen nur m...

Page 76: ...icht Zum Schutz gegen Feuer die Sicherungen nur mit Sicherungen ersetzen deren Typ und Nennwert im Abschnitt Sicherungen ersetzen des Kapitels Wartungsverfahren angegeben sind Attenzione per garantire...

Page 77: ...nd their associated statements can appear on labels affixed to the packaging in which instruments devices and component parts are shipped Safety ground Frame or chassis terminal Fuse Functional ground...

Page 78: ...December 5 2017 715004753 Rev B Page 78 Upper limit of temperature Lower limit of temperature Temperature limitation Symbol Description...

Page 79: ...vironmental specifications for the 2998 PDA detector B 3 Electrical specifications The following table lists the electrical specifications for the 2998 PDA detector Table B 1 Physical specifications A...

Page 80: ...ee 2 A measure of pollution on electrical circuits which may produce a reduction of dielectric strength or surface resistivity Degree 2 refers only to normally nonconductive pollution Occasionally how...

Page 81: ...254 nm Bandwidth 3 6 nm 3 pixel bunch Warm up time 60 minutes Environmental stability 2 C h Flow cell dry analytical 10 mm Sampling rate 2 points s Linearity b 5 at 2 0 AU propylparaben series at 257...

Page 82: ...Wetted materials Standard 316 stainless steel fluoropolymer fused silica PEEK ACQUITY Arc Bio Fluoropolymer fused silica MP35N PEEK titanium a US patent number 5 883 721 Table B 6 2998 PDA detector fl...

Page 83: ...through 0 45 m filters before their use Solvents distilled in glass generally maintain their purity from lot to lot use them to ensure the best possible results C 1 3 Preparation checklist The follow...

Page 84: ...scibility If you are running a high temperature application consider the effect of the higher temperature on solvent solubility Buffers dissolved in water can precipitate when mixed with organic solve...

Page 85: ...7 7 15 4 2 Tetrahydrofuran 0 55 66 0 17 220 4 3 Ethyl acetate 0 47 77 1 19 260 4 3 1 propanol 2 30 97 2 15 210 4 3 2 propanol 2 35 117 7 15 4 4 Methyl acetate 0 45 56 3 15 17 260 4 5 Methyl ethyl keto...

Page 86: ...ibility C 3 Buffered solvents When using a buffer use a good quality reagent and filter it through a 0 45 m filter Do not leave the buffer stored in the system after use Flush all fluid line pathways...

Page 87: ...lvent degassing Mobile phase difficulties account for 70 or more of all liquid chromatographic problems Using degassed solvents is important especially at wavelengths below 220 nm Degassing provides s...

Page 88: ...s section describes the solvent degassing techniques that will help you to attain a stable baseline Degassing your solvent also improves reproducibility and pump performance You can use either of the...

Page 89: ...unit volume of solvent are removed Surface area of the degassing membrane The length of the degassing membrane is fixed in each vacuum chamber To increase the length of membrane you can connect two or...

Page 90: ...sulfide 290 Toluene 285 Ethylene dichloride 230 Xylene 290 Table C 3 Wavelength cutoffs for different mobile phases Mobile Phase UV Cutoff nm Mobile Phase UV Cutoff nm Acetic acid 1 230 Sodium chlori...

Page 91: ...to determine the best wavelength for a particular analysis Table C 4 Electronic absorption bands of representative chromophores Chromophore Chemical Configuration max nm max L m cm max nm max L m cm E...

Page 92: ...centration of the mobile phase affects its absorbance Examples of several mobile phases are given in the table below Tip The absorbances in the table below are based on a 10 mm pathlength Allene C C 4...

Page 93: ...02 0 01 Triethylamine 1 2 33 2 42 2 50 2 45 2 37 1 96 0 50 0 12 0 04 0 01 Buffers and Salts Ammonium acetate 10 mM 1 88 0 94 0 53 0 29 0 15 0 02 0 01 Ammonium bicarbonate 10 mM 0 41 0 10 0 01 0 01 EDT...

Page 94: ...rs PIC reagents PIC A 1 vial L 0 67 0 29 0 13 0 05 0 03 0 02 0 02 0 02 0 02 0 01 PIC B6 1 vial L 2 46 2 50 2 42 2 25 1 83 0 63 0 07 0 01 PIC B6 low UV 1 vial L 0 01 0 01 PIC D4 1 vial L 0 03 0 03 0 03...

Reviews: