Troubleshooting
85
5
5.3 Troubleshooting Chromatographic Problems
is a guide to troubleshooting and correcting chromatographic problems.
Table 5-3 Correcting Chromatographic Problems
Symptom
Possible Cause
Corrective Action
Erratic retention times
Air bubble in pump head
Degas all eluents and prime
the pump (see
Malfunctioning check valves
Replace pump check valves
(see
or
).
Leaking pump seals
Replace pump seals (see
or
Leaks around fittings
Tighten fittings.
Separation chemistry
Check the eluent and column.
Clogged eluent filters
Replace filters.
Increased retention
times
Incorrect flow rate
Change the flow rate.
Incorrect eluent composition
Change eluent composition.
Column heater not turned on
Turn the column heater on.
Incorrect eluent
Use the correct eluent.
Column contaminated
Clean or replace the column.
Incorrect column
Use the correct column.
Air bubble in pump head
Degas all eluents and prime
the pump (see
Clogged eluent filter
Replace the filter.
Reduced retention
times
Incorrect flow rate
Change the flow rate.
Incorrect eluent composition
Change the composition.
High column temperature
Reduce the column
temperature.
Incorrect eluent
Use the correct eluent.
Column contaminated
Clean or replace the column.
Incorrect column
Use the correct column.
Eluent not properly degassed
Degas the eluent.