Troubleshooting Pump Problems
80
5
Pump does not run
Pump not connected to power
source
Ensure the power cable is
properly connected to the
power source and pump.
No power at outlet
Check the outlet by connecting
to another electrical unit
known to be working. If that
unit does not work, relocate
the pump to a functioning elec-
trical outlet.
Blown fuse
Replace fuse.
Pump not delivering
eluent
Draw-off valve open or leaking
Close the draw-off valve. If
eluent still leaks, replace the
valve (see
Pump low-pressure limit set
higher than operating pressure
Set the correct low-pressure
limit.
Flow rate set to zero
Set the desired pump flow
rate.
Pressure transducer out of
adjustment or defective
Contact Waters Technical
Service.
Pump not primed
Section 3.2.1, Priming the
Pump
).
Table 5-1 Correcting Pump Problems (Continued)
Symptom
Possible Cause
Corrective Action