Installing the 1525 Pump
34
2
2.3 Making Electrical Connections
For proper operation, the pump requires:
•
A grounded AC power supply with no abrupt voltage fluctuations
•
A unique IEEE-488 address
•
An IEEE-488 connection to a MassLynx workstation and other IEEE-488 devices
•
An Inject Start signal output connection (if a manual injector is installed)
Required Materials
•
2-meter IEEE-488 cable with a GPIB connector (in Startup Kit)
•
Power cord (in Startup Kit)
•
Small, flat-blade screwdriver (required when connecting the Inject Start signal)
2.3.1 Connecting the Power Supply
The 1525µ and 1525EF Pumps automatically adjust for AC input voltage.
To make the power supply connection:
1.
Insert the 120-V or 240-V power cord into the power cord connection on the rear of
the pump.
2.
Plug the male end of the power cord into a grounded power outlet.
2.3.2 Setting the IEEE-488 Address
Set a unique address for the 1525µ or 1525EF Pump (and for each device connected to
the MassLynx workstation on the IEEE-488 bus). Valid IEEE-488 instrument addresses
are 2 through 29, and are set using the DIP switches on the rear panel.
To set the IEEE-488 address:
1.
Ensure that the power switch is off.
2.
Set the DIP switches on the rear panel of the pump to a unique IEEE-488 address.
Refer to