105/118
24-3. Connection of Analog Output
*NC is blank terminal, but do not use it as relay terminal.
25. BCD Output (Output)
25-1. BCD Output Setting
・
When BCD output is mounted, it can be performed.
・
Setting procedure → "MODE" key → Caution Screen for Starting Setting → Main setting screen →
"PAGE" key → "Output" key.
1) Pattern Select
・
Set a pattern to be set on the set-value select screen.
・
Setting range : 0 to 15
* Select a Pattern Select before changing the set value.
2) Pattern Copy
・
Set the pattern of the copy destination on the set-value select screen.
・
Setting range : 0 to 15
3) Output Channel
・
Set the intended channel through which the indicated value is output on
the select screen.
・
Setting items : Ach, Bch
25-2. Input and Output Specifications
1) Output open collector (NPN type)
・
BCD signal
When logic "1"
transistor is "ON"
・
Polarity (POL) When minus display
transistor is "ON"
・
Overflow signal(OVER) When overflow display transistor is "ON"
・
Print command signal(PC) During (Td × 0.5) transistor is "ON" Td
=
(1/sampling cycle)
2) Equivalent output circuit
・
Open collector output capacity
Voltage MAX. 30V
Current MAX. 15mA
Output saturation voltage above 1.2V when 15mA
Terminal
No.
Direction of
signal
Contents
1
Output
Voltage output
2
*
3
Common (-)
For voltage output
4
*
5
Common (-)
For current output
6
*
7
*
8
*
9
*
10
Output
Current output
1
2
3
4
VOUT
COM
5
6
7
8
9
10
NC
NC
COM
NC
NC
NC
NC
AOUT
Screw terminal for I/O units
Inside of meter
Peripheral equipment
+
COM
D.COM
Output transistor
R
Output
Power supply
5min
φ3.2mm
5.9max
Conforming crimp-type
terminal size