3 3/4 Digits Auto Range DMM with Capacitance and Frequency   

OPERATION MANUAL 

1.  SAFETY RULES 

 

 

This meter is designed and tested in accordance with EN publication 61010-1, pollution 

degree II and installation category (overvoltage category) III 600V. 

 

This meter has been tested according to the following EC Directives 

 

89/336/EEC Electromagnetic Compatibility, EN61326

 

 

73/23/EEC Product safety law of Low Voltage Directive, EN61010-1 

 

This meter is designed to be indoor use at temperature 5

°

C to 40

°

C and altitude up to 

2,000m. 

 

To ensure that the meter is used safely, follow all safety and operating instructions in 

this operation manual.    If the meter is not used as described in this operation manual, 

the safety features of this meter might be impaired. 

 

2.  INTERNATIONAL SYMBOLS 

 

Important information 

 

Diode 

 

see manual 

 

Continuity 

 

AC 

 

Ground 

 

DC 

 

Double insulation 

 

3.  SPECIFICATIONS 

3.1 General Specifications 

Display 

:  3 3/4 digit LCD with max. reading of 3999. 

Polarity 

:  Automatic, (-) negative polarity indication. 

Zero adjustment 

:  Automatic. 

Over range indication 

:  Only the MSD “OL” is displayed. 

Power 

:  9V NEDA 1604, JIS 006P, IEC6F22 battery x 1 

Dimension 

:  194(L) x 93(W) x 39.4(H)mm. 

Net Weight 

:  Approx. 400g. (Including battery). 

 

3.2 Electrical Specifications 

Accuracies are 

±

 (% of r number of least significant digits) at 23

°

±

 5

°

C, <75% RH. 

DC Voltage 

Range  Resolution  Accuracy  Input Impedance  Overload Protection 

400mV 

0.1mV 

±

(1.0%+2) 

10M

 

 

600V DC/AC rms   

(< 30 sec.) 

 

4V 

0.001V 

±

(0.8%+1) 

40V 

0.01V 

400V 

0.1V 

600V 

1V 

DC 

Milli-Voltage  400mV 

0.1mV 

±

(1.0%+2) 

10M

 

600V DC/AC rms   

(< 30 sec.) 

AC Voltage 

Range  Resolution  Accuracy 

Freq.   

Range 

Input 

Impedance 

Overload 

Protection 

400mV* 

0.1mV 

±

(2.0%+5d) 

40~ 

400Hz 

10M

 

600V DC/AC rms 

(< 30 sec.) 

4V 

0.001V 

±

(1.2%+3d) 

40V 

0.01V 

400V 

0.1V 

600V 

1V 

DC Current 

Range  Resolution  Accuracy 

Voltage Drop 

Overload Protection 

400uA 

0.1uA 

±

(1.0% + 5d) 

400mV 

Fuse 500mA/600V 

4000uA 

1uA 

40mA 

0.01mA 

400mA 

0.1mA 

4A 

0.001A 

±

(1.2% + 5d) 

400mV 

Fuse 10A/600V 

10A 

0.01A 

±

(1.5% + 5d) 

200mV 

Fuse 10A/600V 

AC Current 

Range  Resolution  Accuracy 

Voltage 

Drop 

Freq.   

Range 

Overload 

Protection 

400uA 

0.1uA 

±

(1.2% + 5d)  400mV  40-400Hz 

Fuse 

500mA/600V 

4000uA 

1uA 

40mA 

0.01mA 

400mA 

0.1mA 

4A 

0.001A 

±

(1.5% + 5d)  400mV  40-400Hz  Fuse 10A/600V 

10A 

0.01A 

±

(2.0% + 5d)  200mV  40-400Hz  Fuse 10A/600V 

Resistance 

Range  Resolution  Accuracy 

Open circuit 

voltage 

Overload Protection 

400

 

0.1

 

±

(1.0%+2d) 

<700mV 

600V DC/AC rms 

(< 30 sec.) 

4 k

 

0.001 k

 

40 k

 

0.01 k

 

400 k

 

0.1 k

 

4 M

  0.001 M

 

30 M

  0.01 M

 

±

(2.0%+5d) 

Diode Test 

Forward Volt. Drop 

Test Current  Open circuit 

voltage 

Overload Protection 

0-1.000V 

Approx. 

0.6mA 

Max. 1.5V 

600V DC/AC rms 

(< 30 sec.) 

Continuity 

Test 

Test range 

Test Current  Open circuit 

Voltage 

Overload Protection 

Buzzer sounds when 

resistance value 

50

  Approx. 1mA  Approx. 0.5V 

600V DC/AC rms 

(< 30 sec.) 

Capacitance 

Range 

Resolution 

Accuracy 

Overload Protection 

50.00nF 

0.01nF 

±

(3.0%+5d) 

600V DC/AC rms 

(< 30 sec.) 

500.0nF 

0.1nF 

5.000uF 

0.001uF 

50.00uF 

0.01uF 

100uF 

0.1uF(15sec) 

±

(3.5%+5d) 

Frequency 

Range 

Resolution 

Accuracy 

Overload protection 

5Hz 

0.001Hz 

±

(1.5% + 5d) 

Vpp = 600mV 

600V DC/AC rms 

(< 30 sec.) 

50Hz 

0.01Hz 

500Hz 

0.1Hz 

5kHz 

1Hz 

50kHz 

10Hz 

100kHz 

100Hz 

*ACmV only available in manual-range 

 

 

 

4.  PANEL DESCRIPTIONS 

 

1. 

LCD display 

2. 

HOLD key 

3. 

Multifunction key 

4. 

RANGE key 

5. 

Yellow key 

6. 

Multifunction Selector 

7. 

Current input terminal (10A) 

8.

 

Input terminal

 

9.

 

Current input terminal (uA, mA)

 

10.

 

Common input terminal

 

 

 

 

 

5.  OPERATION 
 

!

WARNING

 

1)  When measuring voltage ensure that the instrument is not connected or switched to a 

current, resistance, frequency, duty cycle, capacitance or diode/ continuity check range. 

Always ensure that the correct terminals are used for the type of measurement to be 

made. 

2)  Use extreme care when measuring voltage above 50V, especially from sources where 

high energy exists. 

3)  Avoid making connections to “live” circuits whenever possible. 

4)  Before performing resistance, capacitance, diode or continuity test, ensure that the circuit 

under test is de-energised. 

5)  Always ensure that the correct function and range is selected. If in doubt about the correct 

range, start with the highest and work downwards. 

6)  Extreme care should be taken when using the instrument to conjunction with a current 

transformer connected to the terminals. High voltage may be produced at the terminals if 

an open circuit occurs. 

7)  Ensure that the test leads and prods are in good condition with no damage to the 

insulation. 

8)  Take care not to exceed the overload limits as given in the specifications. 

 

5.1  DC and AC voltage measurement 
1.  Connect the black test lead to the “COM” terminal and red test lead to the “

V

Hz %

” 

terminal. 

2.  Set the multifunction selector to 

  position and connect the test leads across the 

source or load under measurement. 

3.  Using the yellow key to select AC voltage mode. The icon “

~

“  must be shown on the 

display. 

4.  Connect the test leads across the source or load under measurement. 

 

5.2  DC milli-voltage measurement 
1.  Connect the black test lead  to the “COM” terminal and red test lead to the “

V

Hz %

” 

terminal. 

2.  Set the multifunction selector to 

mV

  position and connect the test leads across the 

source or load under measurement. 

3.  Connect the test leads across the source or load under measurement. 

 

5.3  DC and AC current measurement 
1.  Connect the black test lead to the “COM” terminal and red test lead to the “

mA

A

” 

terminal for measurement up to 400mA. 

2.  Set the multifunction selector to desired current range position. 

3.  Connect the test leads in series with the current source to be measured. 

4.  Using the yellow key to select AC current mode. The icon “~“  must be shown on the 

display. 

5.  For current measurement from 400mA  to  10A  (fused) follow generally the above 

procedure but connect the red test lead to “10A” terminal. 

!

CAUTION: Max. input over-load: 600V rms < 10sec. 

 

5.4  Resistance measurement 
1.  Connect the black test lead to the “COM” terminal and red test lead to the “

V

Hz %

” 

terminal.   

2.  Set the multifunction selector to 

  position. 

3.  Connect the test leads across the circuit to be tested. 

!

CAUTION:    Ensure that the circuit to be tested is “dead”. 

Max. input overload: 600V rms < 30sec.           

 

 

5.5  Diode test 
1.  Connect the black test lead to the “COM” terminal and red test lead to the “

V

Hz %

” 

terminal. 

2.  Set the multifunction selector to 

  position. 

3.  Using the yellow key select a 

  function. The icon “

“ must be shown on the display. 

4.  Push on the meter and connect the black and red test leads to the cathode (-) and anode 

(+) ends of the diode to be tested respectively. 

5.  Read the forward voltage drop (junction) value from the display. When the forward biased 

is open, the display will shows overload ‘.OL’ . 

!

CAUTION: Max. input overload: 600V rms < 30sec. 

 

5.6  Continuity test 
1.  Connect the black test lead to the “COM” terminal and red test lead to the “

V

Hz %

” 

terminal. 

2.  Set the multifunction selector to 

  position. 

3.  Using the select key select a    function. The icon “ “ must be shown on the display. 

4.  Connect the test leads across the circuit to be tested, if the resistance less than 50

Ω

buzzer will be activated. 

!

CAUTION: Max. input overload: 600V rms < 30sec. 

 

5.7  Frequency measurement 
1.  Connect the black test lead to the “COM” terminal and red test lead to the “

” 

terminal. 

2.  Set the multifunction selector to 

Hz %

 position   

!

!