50/118
11. Hold Function (Hold)
・
This function is provided to detect one sample with a peak, a valley, a peak valley, a maximal/ minimal,
and an inflection point, holds the indicated value, performs HI/LO limit comparison simultaneously, and
outputs the result.
・
When
「
6. Basic Function Setting (Setup) 6) Input ON Delay
」
is set, it is adopted.
(Input ON Delay is used for STA/STB signal and "HOLD" key.)
* Caution
When STA/STB signal be turned ON and STA/STB signal is turned OFF during Input ON Delay,
Input ON Delay operation is stopped. (When "HOLD" key is turned ON during ON delay, Input ON
Delay operation is also stopped.)
・
"HOLD" key is active in the A channel, B channel, and 2 channel to be set on
「
6. Basic Function Setting
(Setup)
」
. In case of the multi display screen, it is active in both Ach and Bch.
・
"HOLD" key can be used in A channel, B channel (single display screen), and 2 channel (multi-display
screen).
When "HOLD" key is turned ON in 2 channel (multi display screen), Ach and Bch (single display screen)
operate simultaneously. Turning ON "HOLD" key with both STA/STB signals turned ON will not be
accepted. Priority is given to STA/STB signal.
11-1. Hold Function Setting
・
Setting procedure → "MODE" key → Caution Screen for Starting Setting →
Main setting screen → "HOLD" key
Itemized setting screen 1/3
1) Pattern Select
・
Set a pattern to be set on the set-value select screen.
Setting range : 0 to 15
*Select a Pattern Select before changing the set value.
2) Pattern Copy
・
Set the pattern of the copy destination on the set-value select screen.
Setting range : 0 to 15
3) Hold type (Type)
・
Set a hold type among 17 different types on the set-value select screen.
(1) Normal motion (Normal)
(2) Sample hold
(Sample)
(3) Peak hold
(Peak)
(4) Valley hold
(Valley)
(5) Peak valley hold
(Peak Valley)
(6) Area peak hold (Area Peak)
(7) Area valley hold (Area Valley)
(8) Area peak valley hold (Area Peak Valley)
(9) Time peak hold (Time Peak)
(10) Time valley hold (Time Valley)
(11) Time peak valley hold (Time Peak Valley)
(12) Level + time peak hold (Level + Time P)
(13) Level + time valley hold (Level + Time V)
(14) Level + time peak valley hold (Level + Time P-V)
(15) Level + maximal value hold (Maximal Value)
(16) Level + minimal value hold (Minimal Value)
(17) Level + Inflection point hold (Inflection Point)
a) When (1) Normal is selected, "STA" and "STB" signal input or "HOLD" key
are inactive.
b) In case of single or multi display screen, the indicated value of the point
appropriate for the type of hold is held for output.