4.1.2 

 Comparator Data Setting

 

 
 
 
 
 
 
 
 
 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

●Comparative operation type 

In the AM-215A, use the comparator data to select one type of 
comparative operation from the three types below:

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 

4.1.3 

 Scaling Data Setting

 

●Displayed value setting 

The  concept  of  the  scaling  data  and  an  example  setting  of  a 
full-scale value are presented below: 
Displayed value = (a×X) + b 
a = (Displayed full-scale value - Displayed offset value) / (Input 
full-scale value - Input offset value) 
b = Displayed offset value - (Input offset value×a) 
where 
X: Input value; a: Gain; b: Offset

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

●Analog output setting 

The range of displayed values of the AM-215A’s analog output can 
be arbitrarily set, such that it corresponds to the range of 4-20 
mA or 0-10 V. Analog output scaling sets a displayed value to the 
analog output HI that corresponds to 20 mA or 10 V. Likewise, it 
sets a displayed value to the analog output LO that corresponds 
to 4 mA or 0 V. 

The scaling concept is presented below: 
The operation for scaling is the same as that for the displayed 
value setting.

 

 
 
 
 
 
 
 
 
 
 

 
 
 
 

Instruction Manual for MODEL    AM-215A UL                                                                  3/4 

Comparator condition 

HI limit value

HI / GO / LO mode

Measurement value 

Output

OUT1

OUT2

OUT3

HI limit value

Measurement value 

LO limit value ≦

Measurement value 

LO limit value >

OFF

OFF

ON

ON

OFF

OFF

ON

OFF

OFF

Comparator condition 

HH limit value

HH / HI / GO mode

Measurement value 

Output

OUT1

OUT2

OUT3

HI limit value

Measurement value 
Measurement value 

HI limit value

OFF

ON

ON

OFF

OFF

ON

OFF

OFF

ON

Comparator condition 

LO limit value

GO / LO / LL mode

Measurement value 

Output

OUT1

OUT2

OUT3

Measurement value 
Measurement value 

LL limit value

LO limit value

OFF

OFF

ON

ON

OFF

OFF

ON

OFF

ON

(Digital limiter HI  

                 value)

(Inputoffsetvalue)

(Displayedoffset  

                 value)

(Inputfull-scalevalue)

(Displayedfull-scale 

                     value)

  ~

The value to be displayed at the time of FIN signal 
input is set.

  ~

The upper limit value of the displayable range is set.

(Analog output HI 

                 value)

Press twice 

(Decimalpoit)

  ~

Input full-scale signal.

The value to be displayed at the time of OIN signal 
input is set.

Input offset signal.

  ~

  ~

  ~

A displayed value to obtain an output from the full-
scale side of the analog output is set.

  ~

  ~

At sahdira

Flashing decimal indicator is tumed ON/OFF.

*The defaultvalue is OFF (during setting, all of the indicators are 
tumed ON).

If M key is pushed in the state of each menu name or there is no key operation 
during 1 second, it will shift to contents selection of a parameter. Moreover, 
when there is no key operation during 8 seconds at the time of selection of the 
contents of a parameter, it returns to a menu.

(Measurement operation)

Digit shift

Numerical value or option 
change

Back to measurement 
operasion

Default value

The contents of a parameter 

Menu name

(Digital limiter LO  

                 value)

(Analog output LO 
                 value)

*The default value is 9999.

*The default value is 9999 in the 11-13 range, 20.00 in the 2A 
range, and 5.000 in the 1V range.

*The default value is 0.

*The default value is 0 in the 11-13 range, 4.00 in the 2A range, and 
1.000 in the 1V range.

*The default value is 9999.

The lower limit value of the displayable range is set.

*The default value is -9999.

*The default value is 9999.

*A menu is displayed only the option output form 4.

*The default value is 0.

A displayed value to obtain an output from the full-
scale side of the analog output is set.

*          Set up by the key.

Input
voltage

5000

10 V

I n p u t 
v o l t a g e

: 0-10 V

: 0-5000

D i s p l a y e d 
value range

F S C : 5000

F I N : 9999

O F S : 0

O I N

0

0

3000

Displayed 
value range

O u t p u t 
c u r r e n t

: 0-5000

: 4-20 mA

Displayed 
v a l u e 
r a n g e

A O H I : 5000

A O L O : 0

20 mA

5000

4 mA

0

Displayed 
value 
range

Output 
current

Displayed value when the analog output is 20 mA or 10 V

Displayed value when the analog output is 4 mA or 0 V

Caution

When the equipment input range is used, the full scale input values are initially set to 1 V 

 5.000 

and 2 A 

 20.00. The offset input value is 0, so 1 V 

 1.000 and 2 A 

 4.00 are set.

!

(Output 1 Logic)

(Hysteresis 1)

(Check value 1)

(Comparative 

outputtypr)

Normally open

Normally closed

  ~

Check value 1 is set (setting varies depending on the 
COM.T setting.

HI-LO check operation (HI/GO/LO comparative output)

  ~

~            

Hysteresis 1 is set (setting varies depending on the 
COM. Tsetting)

AL1 is turned on by HH.

(AL2 lighting 

           selection)

If M key is pushed in the state of each menu name or there is no key operation 
during 1 second, it will shift to contents selection of a parameter. Moreover, when 
there is no key operation during 8 seconds at the time of selection of the contents 
of a parameter, it returns to a menu.

(Measurement operation)

Digit shift

Numerical value or option 
change

Back to measurement 
operasion

Default value

The contents of a parameter 

Menu name

(Check value 2)

(Hysteresis 2)

(Output 2 Logic)

(Output 3 Logic)

HH check operation (HH/HI/GO comparative output)

LL check operation (GO/LO/LL comparative output)

Check value 2 is set (setting varies depending on the 
COM.T setting.

*The default value is 1000.
*It is set to S-HH at the time of HHHG selection. 

*It is set to S-LO at the time of GLLL selection. 

*The default value is 500.
*It is set to S-HH at the time of HHHG selection. 

*It is set to S-LO at the time of GLLL selection. 

*The default value is 0.
*It is set to H-HH at the time of HHHG selection. 

*It is set to H-LO at the time of GLLL selection. 

~            

Hysteresis 1 is set (setting varies depending on the 
COM. Tsetting)

*The default value is 0.
*It is set to H-LL at the time of HHHG selection. 

*It is set to H-GO at the time of GLLL selection. 

Normally open

Normally closed

Normally open

Normally closed

*It is set to L-HH at the time of HHHG selection. 

*It is set to L-GO at the time of GLLL selection. 

*It is set to L-HI at the time of HHHG selection. 
*It is set to L-LO at the time of GLLL selection. 

*It is set to L-GO at the time of HHHG selection. 
*It is set to L-LL at the time of GLLL selection. 

AL1 is turned on by HI.

AL1 is turned on by GO.

AL1 is turned on by LO.

AL1 is turned on by LL.

AL2 is turned on by HH.

AL2 is turned on by HI.

AL2 is turned on by GO.

AL2 is turned on by LO.

AL2 is turned on by LL.

*The parameter of AL1 and AL2 is set to GO with both sides 
immediately after changing a comparison output type parameter. 

(AL1 lighting 

           selection)

*An output terminal will be out3.

*An output terminal will be out2.

*An output terminal will be out1.

Summary of Contents for AM-215A

Page 1: ...TOR ANALOG OUT or RS485 INPUT EXC POWER c e c e c e AL1 AL2 SET POINT COMPARATOR AM 215A INSTRUCTION MANUAL Caution 1 The equipment may be damaged if the input current or voltage exceeds the maximum...

Page 2: ...y When using a 2 wire sensor terminals 2 11 should be externally connected 4 When no terminals are selected for outputting do not connect anything to vacant terminals AL1 AL2 Condition data A group of...

Page 3: ...et The default value is 9999 The default value is 9999 A menu is displayed only the option output form 4 The default value is 0 A displayed value to obtain an output from the full scale side of the an...

Page 4: ...g temperature and humidity ranges 0 to 50 C 35 to 85 RH no condensation Storage temperature and humidity ranges 20 to 70 C 60 or lower RH no condensation Supply voltage DC 24 V 20 Consumed power Appro...

Page 5: ...inearity compensation of the rectifying point 2 is inputted The input value display value before linearity compensation of the rectifying point 16 is inputted The output value display value before lin...

Page 6: ...ransmission control procedure Signal name used Number of units that can be connected Transmission line length Delimiter Parity bit Start stop Two wire half duplex polling selecting 2400 4800 9600 1920...

Page 7: ...to ON AL1 AL2 AL1 AL2 AL1 AL2 AL1 AL2 Function 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 Char Length 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 Char Length Establishment of...

Page 8: ...0 H I CR LF 10 reading w o decimal point 5 0 0 0 H I CR LF 11 reading w decimal point 9 8 0 0 H I CR LF 10 overrange reading w o decimal point 9 8 0 0 H I CR LF 11 overrange reading w decimal point M...

Page 9: ...3 Response with the status of the hold function being remote controlled D Z R CR LF 3 Response with the status of the digital zero function being remote controlled R L Y CR LF 3 Response with the sta...

Page 10: ...T 1 W 1 CR LF 14 Response with the status of the tracking zero time being 1 and tracking zero width being 1 T R K O N T 9 9 W 9 9 CR LF 16 Response with the status of the tracking zero time being 99 a...

Page 11: ...O side logic function being set to N C N CR LF 1 L L O N O CR LF 8 Response with the status of the LO side logic function being set to N O L L O N C CR LF 8 Response with the status of the LO side log...

Page 12: ...on correction data items being 16 Setting for number of linearization L N O 0 2 CR LF 6 Y E S CR LF 5 correction data items Sets the number of linearization correction data items to 02 L N O 1 6 CR LF...

Page 13: ...unication parameter failure Note Under normal conditions this response is made only once Note ERROR B to ERROR F are available as the response commands for acknowledgment Note If a response to the ERR...

Reviews: