background image

③,④: Output terminals 

· An analog output (4-20 mA or 0-10 V) or an RS485 interface can 

be selected. 

⑦: DZ (Digital Zero) 

The previously displayed value is set to zero. After this setting, 
measurement is performed based on this “zero” point for display. 
When the b.uP is OFF, turning the power OFF will cancel this 
setting. 

* The DZ terminal is enabled by short-circuiting it with the COM 

terminal or setting it to level “0”. It is disabled by 
open-circuiting it or setting it to level “1”. 

 

Level “0”: 0-1.5 V 

Level “1”: 3.5-5 V 

 

Input current: -0.5 mA or below 

⑧, ⑨: Power source terminals (DC POWER) 

· A power source is connected to a power source terminal. The 

AM-215A does not have a power switch. The power is turned ON as 
soon as the power source is connected. 

⑩, ⑪: EXC (sensor power source) 

· An EXC terminal can be used as a sensor power source (24 V DC 

25 mA maximum). 

⑬ - ⑱: Comparative output terminals 

· Relay output 
  Contact capacity:      24 V DC/1 A (resistive load) 
  Mechanical lifespan:   5 million times min. 
  Electrical lifespan:   100,000 times min. (resistive load) 
· Photocoupler output 
  Output capacity:   Voltage: 30 V max., Current: 50 mA max. 
  The maximum output saturation voltage is 1.2 V at 50 mA. 

* Use the relay output and photocoupler output within the range of 
the rated capacity. 

 
 
 
 
 

4  Parameter Setting 

4.1

 Component Name and Function

 

 
 
 
 
 
 

① Main display area:  Displays measured values, as well as menus and 

their contents for parameter setting. 

② Check results display area: Displays the check results (when using the 

GO  output,  neither  the  AL1  nor  the  AL2  is 
flashing). 

③ ENTER switch:       Move from the measurement mode to the parameter 

setting mode (“ENTER” + “Mode”). 

④ Mode switch:        In the parameter setting mode, switch from one item 

to another. 

   

   Move to the shift data setting mode 

 (“Mode” + “Shift”). 

   

Turn the DZ terminal ON/OFF on the front panel 

(“Mode” + “Increment”). 

⑤ Shift switch:      In the parameter setting mode, switch from one 

digit to another. 

   

Move to the shift data setting mode  

(“Mode” + “Shift”). 

⑥ Increment switch:    Select numerical value (“Increment”) and the 

contents for parameter setting. 

   

Turn the DZ terminal ON/OFF on the front panel 
(“Mode” + “Increment”). 

* All parameters can be initialized by switching on a power supply, pushing 

all keys. 

* When the b.uP of the condition data is turned OFF, turning the power 
OFF cancels the digital zero value. 
 

● Parameter group 

The parameters of the AM-215A fall into the five categories below. 
This manual does not explain the settings of the shift data and 
the linearize data. They are explained in a different manual. If 
that manual is needed, please contact our dealer or office. 

 
 
 
 
 
 

4.1.1  Condition Data Setting 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 
 

Instruction Manual for MODEL    AM-215A UL                                                                   2/4 

 

 Caution

(1) Apply the rated voltage at once (gradually increasing the applied voltage may not turn the machine 
ON).
(2) The time interval for turning the power ON/OFF should be at least 10 seconds (turning the power 
ON/OFF within 10 seconds may not turn the machine ON).
(3) T h e   I N P U T   L O ,   E X C   0 V ,   a n d   C O M   t e r m i n a l s   a r e   a t   t h e   s a m e   p o t e n t i a l ,   b u t   t h e y   s h o u l d   b e   w i r e d  
s e p a r a t e l y .   (When using a 2-wire sensor, terminals 2-11 should be externally connected.)
(4) When no terminals are selected for outputting, do not connect anything to vacant terminals.

AL1 AL2

Condition data

A group of parameters that relate to the basic operations and functions of 
the AM-215A, such as sampling rate.

Comparator data

A group of parameters that relate to comparative outputs.

Scaling data

A group of parameters that relate to measurements, such as decimal 
point and displayed value, and input signals.

Shift data

A parameter that relates to the function to forcibly shift a displayed value.

Linearize data

A parameter that performs corrections to determine a straight line 
representing input signals and displayed values.

(Data length)

(Baud rate)

(Digital limiter typ)

(Display blanking)

(Step wide)

(Moving-average 

calculation frequency)

(Average frequency)

(Protect level)

(Parity bit)

(Stop bit)

(Delimiter)

(Equipment ID)

(DZ Backup)

(Linearize)

(TZ correction 

                   time)

(TZ correction 

                  width)

(Delay time when the 

power is turned ON)

All parameters are 
displayed.
Conparator and scaling 
data are displayed.

Only comparator data are 
displayed.
Only pl parameters are 
displayed.

Once (25 times/sec)

OFF(no movingaverage 
calculation)

Normal.

The smallest digit is an 
even numder. 

The smallest digit is 0 or 
5

Brightest

Bright

Dim

Dimmest

Turned off

Retained as a digital limiter value.

Displayed as a value out of the digital limiter value.

9600bps

4800bps

2400bps

38400bps

19200bps

7bit

8bit

Even parity

Odd parity

No parity

1bit

2bit

CR+LF

CR

The ID of RS-485 equipment is set (00 not permitted).

The digital zero value is stored when the power is turned OFF.

The linearize function is not used.

Tracking zero correction time (set value ×times of 
sampling).

Tracking zero correction width (set value ×digit).

No delay time when the power is turned ON.

Digit shift

Numerical value or option 
change

Back to measurement 
operasion

(Measurement operation)

*The default value is 00.

Default value

If M key is pushed in the state of each menu name or there is no key operation 
during 1 second, it will shift to contents selection of a parameter. Moreover, when 
there is no key operation during 8 seconds at the time of selection of the contents 
of a parameter, it returns to a menu.

Twice (12.5 times/sec)

4 times (6.25 times/sec)

8 times (3.125 times/sec)

10 times (2.5 times/sec)

20 times (12.5 times/sec)

40times(0.625 times/sec)

80 times (0.31 times/sec)

100times(0.25 times/sec)

200times(0.13 times/sec)

Twice. 

4 times. 

8 times. 

16 times.

32 times.

The smallest digit is 0 

The digital zero value is not stored when the power is turned 
OFF.

The linearize function is used.

The linearize data is initialized.

*The default value is 00.

*The default value is 00.

No delay time when the power is turned ON is set (set 
value ×second).

The contents of a parameter 

Menu name

*In selection, a menu does not come out of 00 by          . 

*A menu is displayed only the option output form 4.

*A menu is displayed only the option output form 4.

*A menu is displayed only the option output form 4.

*A menu is displayed only the option output form 4.

*A menu is displayed only the option output form 4.

*A menu is displayed only the option output form 4.

Summary of Contents for AM-215A

Page 1: ...TOR ANALOG OUT or RS485 INPUT EXC POWER c e c e c e AL1 AL2 SET POINT COMPARATOR AM 215A INSTRUCTION MANUAL Caution 1 The equipment may be damaged if the input current or voltage exceeds the maximum...

Page 2: ...y When using a 2 wire sensor terminals 2 11 should be externally connected 4 When no terminals are selected for outputting do not connect anything to vacant terminals AL1 AL2 Condition data A group of...

Page 3: ...et The default value is 9999 The default value is 9999 A menu is displayed only the option output form 4 The default value is 0 A displayed value to obtain an output from the full scale side of the an...

Page 4: ...g temperature and humidity ranges 0 to 50 C 35 to 85 RH no condensation Storage temperature and humidity ranges 20 to 70 C 60 or lower RH no condensation Supply voltage DC 24 V 20 Consumed power Appro...

Page 5: ...inearity compensation of the rectifying point 2 is inputted The input value display value before linearity compensation of the rectifying point 16 is inputted The output value display value before lin...

Page 6: ...ransmission control procedure Signal name used Number of units that can be connected Transmission line length Delimiter Parity bit Start stop Two wire half duplex polling selecting 2400 4800 9600 1920...

Page 7: ...to ON AL1 AL2 AL1 AL2 AL1 AL2 AL1 AL2 Function 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 Char Length 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 Char Length Establishment of...

Page 8: ...0 H I CR LF 10 reading w o decimal point 5 0 0 0 H I CR LF 11 reading w decimal point 9 8 0 0 H I CR LF 10 overrange reading w o decimal point 9 8 0 0 H I CR LF 11 overrange reading w decimal point M...

Page 9: ...3 Response with the status of the hold function being remote controlled D Z R CR LF 3 Response with the status of the digital zero function being remote controlled R L Y CR LF 3 Response with the sta...

Page 10: ...T 1 W 1 CR LF 14 Response with the status of the tracking zero time being 1 and tracking zero width being 1 T R K O N T 9 9 W 9 9 CR LF 16 Response with the status of the tracking zero time being 99 a...

Page 11: ...O side logic function being set to N C N CR LF 1 L L O N O CR LF 8 Response with the status of the LO side logic function being set to N O L L O N C CR LF 8 Response with the status of the LO side log...

Page 12: ...on correction data items being 16 Setting for number of linearization L N O 0 2 CR LF 6 Y E S CR LF 5 correction data items Sets the number of linearization correction data items to 02 L N O 1 6 CR LF...

Page 13: ...unication parameter failure Note Under normal conditions this response is made only once Note ERROR B to ERROR F are available as the response commands for acknowledgment Note If a response to the ERR...

Reviews: