Analyzer Initial Start-Up
23
Waltron User Manual 101-046-B.3
9096 Degassed Cation Conductivity Analyzer
4
ANALYZER INITIAL START-UP
Before proceeding with analyzer start-up it is absolutely
necessary to check that all procedures for a proper
installation have been followed. Please verify that all the
suggestions and recommendations have been followed.
4.1
PREPARING THE ANALYZER FOR START-UP
4.1.1
INSTALLING CATION RESIN BOTTLE
Step
Illustration
Description
1
Note the date and time you are
exchanging the resin bottle.
2
Remove cap from new bottle of
resin.
Resin has been pre-rinsed.
3
Inside the cap is a gasket;
remove it from the cap.