User Interface
41
Waltron User Manual 101-046-B.3
9096 Degassed Cation Conductivity Analyzer
5.5
CONFIGURATION
5.5.1
TEMPERATURE COMPENSATION
To configure temperature compensation on the conductivity measurement depending on
chemicals used follow the procedure below.
Action
Display
Configuration of the
temperature
compensation requires
log in access of at least
the Service Level.
Access the
Configuration Menu by
pressing the bottom left
hand corner of the main
display screen to open
the main device menu
and selecting
Configuration.
From the Configuration
menu select Analog
Inputs > Universal Input
1 > Configuration.
Press the yellow box for
Operation mode and
select Conductivity
Measurement from the
list of options.